TROUBLESHOOTING
If any abnormality arises during use, please check the following points
Note:
If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair
the unit by yourself. Contact information for HoMedics Consumer
Relations Department can be found on the warranty page.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn on
when the
START/STOP
button
is pushed.
Batteries have
run down.
Replace them with
four new AAA
alkaline batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries
in the correct positions.
EE measurement error
symbol shown on dis-
play or the blood pres-
sure value is displayed
excessively low
(or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly
so that it is positioned
correctly.
Did you talk or move
during measurement?
Measure again. Keep
arm steady during
measurement.
Shaking of the arm with
the cuff on.
Er error symbol shown
on display - measure-
ment result exceeds
measurement range.
Did you move during
measurement?
Shaking of the arm with
the cuff on
Did you move during
measurement?
Shaking of the arm with
the cuff on
E1 error symbol shown
on display
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
Measure again.
E2 error symbol shown
on display
Inflation pressure
exceeding 300 mmHg.
Switch the unit off, then
measure again.
E3 error symbol shown
on display
Data error.
Remove and reload the
batteries.
CA Appears on
display
Recalibration alert.
Refer to page 25 for
more information.
Power Source:
Four 1.5V DC (AAA) batteries or
optional 12V DC AC adapter (not included)
Measurement
Method:
Oscillometric
Measurement Range:
Pressure: 40~250 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
48 memories for each User (96 total)
Auto-shut-off:
1 minute after last button operation
Operation
Environment:
Temperature: 50°F~104°F (10°C ~40°C)
Humidity: 40~85% RH max.
Storage Environment:
Temperature: 14°F~140°F (-10°C~60°C)
Humidity: 10~90% RH max.
Weight:
1.0 lb (462 g) (without batteries)
Arm Circumference:
Standard size cuff: 9.5"–14.2" (24-36 cm)
Large size cuff:13.4"–18.1" (34-46 cm)
Dimensions:
5.6"(L) x 3.7"(W) x 2.8"(H)
142mm(L) x 95mm(W) x 71mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
SPECIFICATIONS
Note:
These specifications are subject to change without notice.