Two 1.5V DC (AAA) Alkaline batteries
M
Me
ea
as
su
urre
em
me
en
ntt
M
Me
etth
ho
od
d::
Oscillometric
M
Me
ea
as
su
urre
em
me
en
ntt
R
Ra
an
ng
ge
e::
Pressure: 40~250 millimeters Mercury
(mmHg)
Pulse: 40~199 beats/minute
A
Ac
cc
cu
urra
ac
cy
y::
Pressure: ±3 mmHg
Pulse: ±5% of reading
P
Prre
es
ss
su
urre
e S
Se
en
ns
so
orr::
Semi-conductor
IIn
nfflla
attiio
on
n::
Pump driven
D
De
efflla
attiio
on
n::
Automatic Pressure release valve
M
Me
em
mo
orry
y C
Ca
ap
pa
ac
ciitty
y::
30 memories for each User (60 total)
A
Au
utto
o--s
sh
hu
utt--o
offff::
1 minute after last button operation
O
Op
pe
erra
attiio
on
n
E
En
nv
viirro
on
nm
me
en
ntt::
Temperature: 50°F~104°F (10°C~40°C)
Humidity: 40%~85% RH max.
S
Stto
orra
ag
ge
e
E
En
nv
viirro
on
nm
me
en
ntt::
Temperature: 14°F~140°F (-10°C~60°C)
Humidity: 10%~90% RH max.
W
We
eiig
gh
htt::
0.25 lbs (without batteries)
W
Wrriis
stt
C
Ciirrc
cu
um
mffe
erre
en
nc
ce
e::
5.3"- 8.6" (13.5 - 22cm)
D
Diim
me
en
ns
siio
on
ns
s::
3.25"(L) x 2.75"(W) x 2.60"(H)
82mm(L) x 69mm(W) x 66mm(H)
Type BF
Device and cuff are designed to
provide special protection against
electrical shocks.
SPECIFICATIONS
Note:
These specifications are subject to change
without notice.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn on
when the
START/STOP
button
is pushed.
Batteries have
run down.
Replace them
with two new AAA
alkaline batteries.
Battery polarities have
been positioned
incorrectly.
Re-insert the
batteries in the
correct positions.
EE measurement
error symbol shown
on display or the
blood pressure value
is displayed exces-
sively low
(or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly
so that it is positioned
correctly.
Did you talk or move
during
measurement?
Measure again.
Keep wrist steady
during measurement.
Shaking of the wrist
with the cuff on.
E1 error symbol
shown on display
Air circuit
abnormality.
Measure again.
E2 error symbol
shown on display
Inflation pressure
exceeding 300 mmHg.
Switch the unit off,
then measure again.
E3 error symbol
shown on display
Data error.
Remove and reload the
batteries.
“
--” appears on
display
Recalibration alert.
Refer to page 28 for
more information.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
Note:
If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page (See page 35).