background image

14

15

• When taking the temperature of an infant younger than three months.
• When taking the temperature of children under the age of

three who have special conditions, such as compromised
immune systems, and for whom the absence or presence
of a fever is critical.

• When you are first learning how to use this ear thermometer

and until you are obtaining consistent readings.

Note:

Slight temperature variations may be observed (+/-

0.3 – 0.5°F or +/- 0.2 – 0.3°C) in consecutive
measurements. This is common to all types of thermometers.

• Always take the temperature in the same ear, as readings

may differ from one ear to another.

• The ear must be free from excessive earwax buildup or other

obstructions in order to take an accurate reading. Note that
you should never attempt to clean inside ears, however, as
this may result in damage to the eardrum or surrounding
tissues. Remove only excess earwax that can be reached
with a washcloth. Consult a physician if you suspect excess
earwax is present within the ear.

• Ear temperatures may be influenced by external factors, including:

• An ear that has been rested upon or covered
• General exposure to extreme hot or cold temperatures
• Recent bathing or swimming
• Use of hearing aids or ear plugs 
It is recommended that measurement occur at least 20 minutes
after these circumstances are no longer present.

• This ear thermometer is for use in human ears only. Do not

use for oral, rectal, or axillary temperature measurement.

SPECIAL SITUATIONS IN WHICH THE EAR
THERMOMETER SHOULD NOT BE USED

• Pain of any kind is noted during use.
• Blood or drainage is present in the external ear canal.
• Deformities of the ear canal or face exist that prohibit the

thermometer from being properly inserted into the ear canal.

• Ear medication or ear drops have been placed in the ear canal.
• Ear infection of any kind is present.
• Individual has been bathing or swimming, as this may lead to

injury of ear tissue, as well as inaccurate readings.

CARE, MAINTENANCE & CLEANING

• Clean the unit carefully with a soft, dry cloth. Do not use any

gas, strong detergent, or solvents to clean the device.

• Clean the waterproof probe with a cotton swab moistened with alcohol. 
• Never attempt to insert any object into the probe tip.
• Do not immerse the unit in water as this will result in damage 

to the unit. Waterproof portion of unit is limited to probe tip.

• Do not subject the ear thermometer to extreme temperatures,

humidity, moisture, and direct sunlight.

• Do not disassemble the ear thermometer. If in need of repair,

refer to the warranty section of this manual.

• Do not subject the ear thermometer to extreme shocks

(do not drop on floor).

• When not in use, keep thermometer in its storage case and

out of the reach of children.

• The thermometer is initially calibrated at the time of manufacture.  If this 

thermometer is used according to the use instructions, periodic re-
adjustment is not required.  If at any time you question the accuracy of

Содержание TE-100

Страница 1: ...DA SERÁ LA ÚNICA Y EXCLUSIVA GARANTÍA NO HABRÁ NINGUNA OTRA GARANTÍA EXPRESA NI IMPLÍCITA INCLUYENDO NINGUNA GARANTÍA DE COMERCIALIZACIÓN O APTITUD NI NINGUNA OTRA OBLIGACIÓN POR PARTE DE LA COMPAÑÍA CON RESPECTO A PRODUCTOS CUBIERTOS POR ESTA GARANTÍA HOMEDICS NO TENDRÁ RESPONSABILIDAD ALGUNA POR NINGÚN DAÑO INCIDENTAL CONSECUENTE NI ESPECIAL BAJO NINGÚN CONCEPTO ESTA GARANTÍA REQUERIRÁ MÁS QUE L...

Страница 2: ... taking a temperature Thermometer should be used in a dry ear canal Gentle force should be used when inserting the probe into the ear Excessive force may injure the ear canal or eardrum Do not expose the monitor to extreme temperature very high humidity or direct sunlight Handle product with care and avoid dropping Do not immerse the unit in water or other liquids Do not disassemble or attempt to ...

Страница 3: ...he average normal body temperature is 98 6 F 37 C Fever is defined as a body temperature elevated over an individual s normal body temperature It is important to determine each person s normal body temperature in order to properly assess whether or not a fever is actually present This range can easily be determined by tracking temperatures on a consistent basis using the same ear and measuring at ...

Страница 4: ...der to determine when a fever is present Classification of a temperature as a fever also varies by age This product is not intended as a substitute for consultation with a physician Contact your physician if you feel an illness is present with or without a fever A physician should be consulted if any of the following conditions are present Vomiting diarrhea changes in breathing activity or appetit...

Страница 5: ...release the F C button located behind the battery compartment cover while the unit is on by using a small flat instrument Once the desired scale is selected press the Scan button to take a new reading or allow the unit to turn off automatically and then turn it back on to view the converted temperature stored in memory Figure 3 Low battery warning c c Indicates a memory reading MEMORY RECALL This ...

Страница 6: ...eter outside of the ear An arrow symbol will appear on the display Figure 3 5 Insert the probe tip of the ear thermometer into the ear canal within two seconds Figure 4 It is important that the probe tip be well positioned within the ear canal Measurement time for the actual reading is approximately one second 6 A beep sound will indicate the end of the measurement Results of the temperature measu...

Страница 7: ...ically turn off after 30 seconds of non use IMPORTANT NOTES REGARDING TEMPERATURE MEASUREMENT The measurement results that users receive are for reference only If users have any health concerns please consult a physician It is not necessary to perform an ear tug or pull on the ear before taking a temperature measurement For additional measurements users should wait at least 5 seconds between readi...

Страница 8: ...r use in human ears only Do not use for oral rectal or axillary temperature measurement SPECIAL SITUATIONS IN WHICH THE EAR THERMOMETER SHOULD NOT BE USED Pain of any kind is noted during use Blood or drainage is present in the external ear canal Deformities of the ear canal or face exist that prohibit the thermometer from being properly inserted into the ear canal Ear medication or ear drops have...

Страница 9: ...tery warning Battery power is low Replace battery see Inserting and Replacing Batteries Low battery warning with Lo symbol Battery power is too low for the thermometer to function Replace battery immediately see Inserting and Replacing Batteries PRODUCT SPECIFICATIONS Model Number Model Number TE 100 Power Source DC 3 Volt lithium battery CR2032 Battery Life 1 700 Measurements Sensor Thermopile Te...

Страница 10: ...HALL BE THE SOLE AND EXCLUSIVE WARRANTY THERE SHALL BE NO OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS OR ANY OTHER OBLIGATION ON THE PART OF THE COMPANY WITH RESPECT TO PRODUCTS COVERED BY THIS WARRANTY HOMEDICS SHALL HAVE NO LIABILITY FOR ANY INCIDENTAL CONSEQUENTIAL OR SPECIAL DAMAGES IN NO EVENT SHALL THIS WARRANTY REQUIRE MORE THAN THE REPAI...

Страница 11: ...guientes de este manual Notas importantes sobre la medida de la temperatura pg 30 Situaciones especiales en las cuales no debe usarse en termómetro auricular pg 32 Cuidado mantenimiento y limpieza pg 33 20 21 IMPORTANTES INSTRUCCIONES DESEGURIDAD Cuando utiliza su termómetro auricular siempre se deben seguir ciertas precauciones básicas Por favor lea y aténgase a todas las instrucciones y adverten...

Страница 12: ...aya necesariamente fiebre Termómetro Auricolar Deluxe Lectura en Un Segundo Cómo funciona este artículo Este termómetro auricular es un dispositivo delicado y de alta tecnología diseñado para medir la temperatura corpórea Esta unidad proporciona lecturas de temperatura precisas y confiables registrando el calor infrarrojo del tímpano y del tejido que lo rodea Este termómetro escanéa 512 veces el t...

Страница 13: ...r así determinar cuando hay fiebre La clasificación de la temperatura como fiebre también varía con la edad Este producto no es entendido como un sustituto de una consulta médica Si cree que está enfermo con o sin fiebre consulte a su médico Es necesario consultar al médico si se presenta cualquiera de las siguientes condiciones vómito diarrea cambios en la respiración actividad o apetito irritabi...

Страница 14: ...ome una medida La batería debe reemplazarse cuando aparece el símbolo de batería baja en la pantalla Fig 3 Para que el termómetro auricular funcione es necesario reemplazar la batería cuando aparece el símbolo de batería baja con el símbolo Lo ATENCIÓN No deseche las baterías de litio usadas en la basura Por favor llévelas a cualquier punto de recolección oficial Cuando no se use la pantalla para ...

Страница 15: ...te termómetro es de 50 F 10 C a 122 F 50 C Si la medida queda debajo de este rango la unidad indica Lo Fig 7 Si queda arriba de este rango la unidad indica Hi Fig 8 28 29 Figura 2 Indica la lectura de la medida Figura 3 Indica que se está tomando la medida Figura 4 3 Cuando el termómetro auricular está listo la pantalla indicará los resultados de la lectura más reciente Ahora la unidad está lista ...

Страница 16: ...iebre es crítica Cuando usted aprende por primera vez el uso de este termómetro auricular y hasta que logra lecturas consistentes ATENCIÓN Es posible observar pequeñas variaciones de temperatura 0 3 0 5 F ó 0 2 0 3 C en medidas consecutivas Esto es común en todo tipo de termómetro Tome la temperatura siempre en el mismo oído puesto que la lectura puede variar de un oído a otro La unidad indica Err...

Страница 17: ...riódica Si en cualquier momento dudase de la exactitude de sus mediciones póngase en contacto con HoMedics Consumer Relations al 1 800 466 3342 Para tomar una lectura correcta el oído debe estar libre de depósitos excesivos de cera y otras obstrucciones Sin embargo observe que usted no debería tratar nunca de limpiar el interior del oído pues esto podría resultar en daños al tímpano o al tejido qu...

Страница 18: ...a por debajo del rango de medida debajo de 93 2 F 34 C Rango Operacional El rango de medida es de 93 2 F 34 C a 109 4 F 43 C Si tiene o sospecha tener cualquier problema de salud póngase en contacto con su médico La sonda no está posicionada correctamente en el oído Asegúrese de que la sonda esté posicionada correctamente en la apertura del canal auricular La sonda impermeable está sucia Limpie la...

Отзывы: