Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The [BPW-1005] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [BPW-1005] should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz to
2.5 GHz
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the
BPW-1005, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2 √
—
P
80 MHz to 800 MHz
d = 2.3 √
—
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the [BPW-1005] is used exceeds
the applicable RF compliance level above, the [BPW-1005] should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
[BPW-1005].
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Содержание BPW-1005
Страница 1: ...BPW 1005 EU Wrist Blood Pressure Monitor Instruction Manual...
Страница 119: ...119 GB...
Страница 120: ...120 GB...
Страница 121: ...121 GB...
Страница 122: ...122 GB...
Страница 123: ...123 GB...
Страница 124: ...124 GB...
Страница 125: ...125 GB...
Страница 126: ...126 GB...
Страница 127: ...127 GB...
Страница 128: ...128 GB...
Страница 129: ...129 GB...
Страница 130: ...130 GB...
Страница 131: ...131 GB...