6.0 MULTI-CHANNEL INSTALLATION
Prior to use of an HS9000B Series Multi-Channel Synthesizer users will need to be setup with the
basic hardware and software.
6.1 HARDWARE INSTALLATION
Prior to initializing the synthesizer, connect the power cord to an active AC power supply. The
instrument is shipped with the appropriate power cord for the final destination country/region. The
master power switch located at the right side of the front panel is equipped with a blue indicator
light which illuminates when the AC power is active.
NOTE:
If the power light is not illuminated while the front panel switch is in the “ON”
position, verify that there is power at the AC outlet/supply and that the fuse has not blown.
A fuse is located in the service tray on the power cord receptacle (rear panel). A spare fuse
is provided inside the service tray.
6.2 HS9000B SERIES APPLICATION GUI
The Holzworth HS9000B Series GUI can be run on any Windows PC, no software installation is
required. The application GUI is contained on the USB drive that was included with the synthesizer.
Simply double click the executable file to launch the GUI. If the USB drive was lost another can be
mailed or software can be downloaded using the URL below.
HS9000B Application GUI:
http://www.holzworth.com/software/Synthesizers/HS9000B/HS9000B.zip