3Dimensions System User Guide
Chapter 2: General Information
MAN-05085-002 Revision 002
Page 17
DRAFT
Preview Copy-Generated May 30, 2018
Caution:
The display is calibrated for compliance to DICOM standards. Do not make any
brightness or contrast adjustments to the display.
Caution:
Use the least possible amount of cleaning fluids. The fluids must not flow or run.
Caution:
To prevent damage to the electronic components, do not use disinfectant sprays on
the system.
Note:
Hologic does not supply the Gantry power cable for some countries. If the power cable
is not supplied, the installed cable must meet the following requirements and all local
codes that apply: 3 conductor, 8 AWG (10 mm
2
) copper not more than 25 feet (7.62
meters) in length.
2.4
Emergency Off Switches
The Emergency Off switches remove the power from the Gantry and the Acquisition
Workstation Lift Mechanism (on the workstations that include that option). Do not
routinely use the Emergency Off switches to turn off the system.
Figure 3: Emergency Off Switch
Functionality
There are three Emergency Off switches, one on each
side of the Gantry and one on the Acquisition
Workstation.
1.
To turn Off the Gantry and disable the Acquisition
Workstation Lift Mechanism, press any of the
Emergency Off switches.
2.
To reset the Emergency Off switch, turn clockwise
approximately one-quarter turn until the switch
pops back out.