background image

 

(2)

 

Press  button 

/

  to  Switch 

/

  or  (number  6 

key). 

Note: 

a

 

If current is in wind flow measuring mode (on the left 
top of the LCD will show "FLOW", you can turn it into 
wind  velocity  measuring  mode  by  pressing  the 
"VEL/FLOW' key, vice versa. 

b

 

The defaulted model is last time you has setting 

 
Measurement of Wind flow 

 

Press the “VEL/FLOW” key to get into “FLOW” mode at this 

time the LCD screen as shown in figure 4 : 

 
 

 

Figure 4 

(if  the  area  value  has  been  set  before  the  last  setting  will  be 
shown) 

 

Select your desired unit of wind flow and duct area: 
a

 

Press the “UNIT” key, the wind flow unit will convert 
between CMM and CFM   

b

 

Press  the  “UNIT”  key,  the  area  unit  will  convert 
between “M

2

” and “FT

2

”.     

c

 

Area unit: "M

2

 " and "FT

2

" will convert according with 

the conversion of the wind flow unit. 
If wind flow unit is CMM, the area unit will be “M

2

”; 

If wind flow unit is CFM. the area unit will be “FT

2

”. 

 

Press  the  “AREA”  key,  the  numbers  in  the  middle  of  the 
LCD will be disappeared at this time press numeric keys to 
enter a new duct area, such as1.2, then press the “ENTER” 
key to confirm. At this time  the LCD screen as shown in 
figure5: 

 

 

 

 

                                  Figure 5

 

 

Place  the  vane  in  the  duct  area,  take  measurement  of 
current wind flow value right away. As shown in figure 6 : 

 
   
 
 
 
 
 

Figure 6 

 

Wind Flow Calculate Formula: Flow =velocity *(Free Area) 

 

Testing  Wind  velocity  :  The  Bar  graphic  (showing  on  the 
middle of LCD )was subject to the increasing wind Flow/ 
Velocity 

Note: without enter the duct area will lead to failing in taking 

measurement of the wind flow. 
If wind flow is larger than 9999, the LCD screen will show 
x10 or x100, and the measured value is also x10 or x100. 

 

Measurement of 2/3 V max Wind Flow 

 

Press the “VEL/FLOW” key to entre "FLOW" mode, the LCD 

screen shows in figure7: 

 

 
 
 
 

   

Figure 7 

 

Select  your  desired  unit  by  pressing  the  “UNIT”  key,  for 
example: select CMM for wind flow unit, “M

2

” for area unit 

at this time the LCD screen shows as above diagram. 

 

Press the "AREA " key, Letter on the LCD will disappeared 

at  this  time  you  can  input  duct  area,  such  as  1.2,  press 
“ENTER”  key  confirm  press  again  the  “OPTION”  key  to 
select “2/3 V max”, the LCD screen as shown in figure 8 

 
 

 
 
 
 

Figure 8 

 

Aim the vane of the attached unit at duct area properly, take 
measurement  of  2/3  V  max  wind  flow  right  away,  press 
again the “OPTION” key exit to measurement of 2/3 V max 
wind flow. 

 

2/3  Max  Wind  flow  calculate  formula:  FLOW  =2/3×Max 

wind Velocity×duct Area. 

 
MEASUREMENT OF AVERAGE WIND FLOW 

 

Press the “VEL/FLOW” key to get into “FLOW” mode, the 

LCD screen shows as in figure 9: 

 

 
 

 

 

Figure 9 

 

Select your desired unit by pressing the “UNIT” key, for 

example: select CMM for wind flow unit, “M2” for area unit, 
at this time the LCD screen shows as above diagram 

 

  press  the  “OPTION”  key  to  select  “AVERAGE”,  the  LCD 

shows as in figure10: 

 
 

 

 

 

                                        Figure10

 

Aim the vane at certain point of duct area properly, press the 
“NEXT” key, In the LCD of the upper part of the serial number, in 
the meantime you can take measurement of the first group wind 
flow values. As shown in figure 11:

 

 

 

 

Figure 11 

Then select another test point, press the “NEXT” key to take 
measurement of the second group average wind flow values. 
Repeat above steps to measure as many as 12 groups average 
wind flow values. As shown in figure 12: 

 

 

 

Figure 12 

Once again press the "OPTION key exit to measurement of 
average wind flow. 

 

Average Wind flow calculate formula 
Flow =l/N  ∑(Velocity)

×

(free Area) 

Note 

The average value only displays by pressing NEXT key, If 
there  is  wind  in  the  duct  area,  then  the  LCD  Screen  will 
shows the measured Average Wind flow. 

 

Max and min measurement     

 

In  wind  flow  and  velocity  measuring  process,  press  the 

“MAX/MIN”  key  to  obtain  the  maximum  and  minimum 
values, press again to exit. For example: 

(1)

 

When measuring maximum value of wind velocity,    a 

“MAX” will be shown on the upper LCD screen, as shown in 
figure 13 

                     

Figure 13

 

(2)

 

When  measuring  minimum  value  of  wind  velocity,  a 
“MIN”  will  be  shown  on  the  upper  LCD  screen,  as 
shown in figure 14: 

Содержание HP-846A

Страница 1: ...easurement of 2 3V max and average wind flow Data holding storing and deleting function Low battery indicating function Auto power off function Button without any operation and the fan static after 5 minutes will be turned off Memory of 300 records Backlight function Pressing key audio alert Large LCD display Wind Handle Can Elongate DIAGRAM OF THE UNIT 1 LCD display 2 Max Min value switch 3 Power...

Страница 2: ... 0 999900 0 001 100 0 001 9999 2 Unit Conversation m s Ft min Knots Km h Mph m s 1 196 87 1 944 3 60 2 24 Ft min 0 00508 1 0 00987 0 01829 0 01138 Knots 0 5144 101 27 1 1 8519 1 1523 Km h 0 2778 54 69 0 54 1 0 6222 Mph 0 4464 87 89 0 8679 1 6071 1 4 Air Temperature Range Unit Scale Resolution Accuracy 0 0 45 0 0 1 1 0 32 0 113 0 0 1 1 8 5 Operation Conditions Temperature Humidity Host 0 50 32 122 ...

Страница 3: ...his time the LCD screen shows as above diagram Press the AREA key Letter on the LCD will disappeared at this time you can input duct area such as 1 2 press ENTER key confirm press again the OPTION key to select 2 3 V max the LCD screen as shown in figure 8 Figure 8 Aim the vane of the attached unit at duct area properly take measurement of 2 3 V max wind flow right away press again the OPTION key ...

Страница 4: ...is block or not connected 3 If the unit can not read the wind temperature value properly please check if the heat resistor is fall off or damaged by manual 4 If the unit can not read data properly please check it is operated under the rule temperature and humidity situation 5 when Measure the wind velocity the temperature display line don t show the number for temperature overflow or probe fault W...

Отзывы: