6-10 Battery

 

range 

Load resistance

 

battery by test is good

 

1.5V 

36Ω 

more than 33mA. 

9V 

360Ω 

more than 18mA. 

12V 

450Ω 

more than 14mA. 

 
7. OPERATING INSTRUCTIONS 
7-1 Attention before operation   

7-1-1 Check 9V battery. If the battery voltage is less than 7V, display will show 

”,  the  battery  should  be  replaced  at  this  time  to  ensure  measuring 

precision. 
7-1-2 Pay attention to the “

” besides the input jack which shows that the 

input voltage or current should be within the specified value. 
7-1-3 The range switch should be positioned to desired range for measurement 
before operation. 

7-2 Measuring DC Voltage 

7-2-1 Connect the black test lead to COM jack and the red to V

Ω

mA jack. 

7-2-2 Set the rotary switch at the desired V

 

range position. 

7-2-3 Connect test leads across the source or load under measurement. 
7-2-4 You can get reading from LCD. The polarity of the red lead connection 
will be indicated along with the voltage value. 

NOTE:

   

1.    When  the  value  scale  to  be  measured  is  unknown  beforehand,  set  the 
range selector at the highest position.

 

2.

 

When  only  the  figure‟1‟  or  „-1‟  is  displayed,  it  indicates  over-range 

situation and the higher range has to be selected.   
3.

 

” means you can‟t input the voltage more than 600V, it‟s possible to 

show higher voltage, but it may destroy the inner circuit or pose a shock.   
4.

 

Be cautious against shock when measuring high Voltage.   

7-3 Measuring AC Voltage 

7-3-1 Connect the black test lead to COM jack and the red to V

Ω

mA jack. 

7-3-2 Set the rotary switch at the desired V~

 

range position. 

7-3-3 Connect test leads across the source or load under measurement. 
7-3-4 You can get reading from LCD.   

NOTE:

   

1.    When  the  value  scale  to  be  measured  is  unknown  beforehand,  set  the 
range selector at the highest position.

 

2.

 

When  only  the  figure‟1‟  or  „-1‟  is  displayed,  it  indicates  over-range 

situation and the higher range has to be selected.   
3.

 

” means you can‟t input the voltage more than 600V, it‟s possible to 

show higher voltage, but it may destroy the inner circuit or pose a shock.   
4.    Be cautious against shock when measuring high Voltage. 

7-4    Measuring DC/AC Current 

7-4-1 Connect the black test lead to COM jack and the red to the V

Ω

mA jack 

for a maximum 200mA current, 
7-4-2 Set the rotary switch at the desired A

/A

~

 

range position. 

7-4-3 Connect test leads in series with the load under measurement. 
7-4-4 You can get reading from LCD. The polarity of the red lead connection 

will be indicated along with the current value. 

NOTE: 

1.

 

When the value scale to be measured is unknown beforehand, set the range 

selector at the highest position.   
2.

 

When only the figure‟1‟ or „-1‟ is displayed, it indicates over-range situation 

and the higher range has to be selected. 

3.

 

”  means  the  socket  mA‟s  maximum  current  is  200mA  and  10A‟s 

maximum current is 10A, over current will destroy the fuse.    Since 10A is not 
fused, the measuring time should be less than 1 second to prevent precision 
from affecting by circuit heating.

 

7-5    Measuring Resistance 

7-5-1 Connect the black test lead to COM jack and the red to V

Ω

mA jack. 

7-5-2 Set the rotary switch at the desired 

Ω

range position. 

7-5-3 Connect test leads across the resistance under measurement. 
7-5-4 You can get reading from LCD.   

NOTE: 

1.

 

When only the figure‟1‟ or „-1‟ is displayed, it indicates over-range situation 

and the higher range has to be selected.   
2.

 

For measuring resistance above 1M

Ω

, the mete may take a few seconds to 

get stable reading. 
3.

 

When  the  input  is  not  connected,  i.e.  at  open  circuit,  the  figure  „1‟ will  be 

displayed for the over-range condition. 
4.

 

When  checking  in-circuit  resistance,  be  sure  the  circuit  under  test  has  all 

power removed and that all capacitors have been discharged fully. 
5.

 

the  value  scale  to  be  measured  is  unknown  beforehand,  set  the  range 

selector at the highest position. 

7-6    Transistor Testing 

7-6-1 Set the rotary switch at ‟hFE‟ position. 
7-6-2 Determine whether the transistor under testing is NPN or PNP and locate 
the emitter, base and collector leads. Insert the leads into proper holes of hFE 
socket on the front panel. 
7-6-3 Read the approximate hFE value at the testing condition of base current   
Ib10uA and Vce 3V. 

7-7    Diode Testing   

7-7-1 Connect the black test lead to COM jack and the red to V

Ω

mA jack. (the 

polarity of red lead is „+‟. 
7-7-2 Set the rotary switch at the 

  F range position. 

7-7-3 Connect the red lead to the anode and the black lead to 
the cathode of the diode under testing. 
7-7-4 You can get a reading from LCD. 

NOTE: 

1. The meter will show approximate forward voltage drop of the diode. 
2. If the lead connections is reversed, only „1‟ will be displayed. 

7-8    Continuity Testing 

7-8-1 Connect the black test lead to COM jack and the red to V

Ω

mA jack. 

7-8-2 Set the rotary switch at the 

  range position. 

7-8-3 Connect test leads across two points of the circuit under testing.   
7-8-4 If continuity exists (i.e. resistance less than about 50

Ω

),    built-in buzzer 

will sound. 

NOTE: 

If the input open circuit, the figure „1‟ will be displayed.   

7-9    Measuring Capacitor 

7-9-1 Connect the black test lead to Cx- jack and the red to Cx+ jack. 
7-9-2 Set the rotary switch at the desired 20uF/200nF range position. 
7-9-3 Before inserting capacitor under measurement into capacitance testing 
socket, be sure that the capacitor has been discharged fully.   
7-9-4 You can get reading from LCD. 

 

7-10    Measuring

 

  Battery                                                                               

  Load resistance    1.5V range: 36Ω      9V range: 360Ω    12V range: 450Ω   
7-10-1 The 1.5V battery by test is good when the display is more than 33mA.   
7-10-2 The 9V battery by test is good when the display is    more than 18mA. 
7-10-3 The 12V battery by test is good when the display is    more than 14mA. 

8. Maintenance 

8-1 Before attempting to remove the battery door or open the case, be sure 
that  test  leads  have  been  disconnected  from  measurement  circuit  top  avoid 
electric shock hazard. 
8-2  To  avoid  electrical  shock,  remove  test  leads  from  measurement  circuits 
before replacing the fuse. For protection against fire, replace fuses only with 
specified ratings: F-200mA/250V fuse. 
8-3 Your must replace the test leads if the lead is exposed, and should adopt 
the leads with the same specifications as origin. 
8-4  Use  only  moist  fabric  or  small  amount  of  detergent  but  not  chemical 
solution for cleaning. 
8-5 do not use the meter before the back cover is properly closed and screw 
secured.    Upon  any  abnormality,  stop  operation  immediately  and  send  the 
meter for maintenance. 
8-6 Please take out the battery when not using for a long time. 

9. Accessories 

[1] Test Leads: electric rating 600V 10A 
[2] Fuse: F-200mA/250V 
[3] Operator‟s Manual 
 

Above picture and content just for your reference. Please be subject to 
the  actual  products  if  anything  different  or  updated.  Please  pardon  for 
not informing in advance. 