Remove the water
Problem
Possible Course
Remedy
Therapy pressure not steady
Check of bends
Measuring tube bent
Water in the measuring tube
Remove the water
Device defective
Notify the service
Common cold
Replace the mask if its material
has become brittle
Increase the air humidity in the
room or set the humidifier to a
higher level
Running nose, urge to sneeze
Reaction to the air flow
Cold nose
Room temperature too low
Increase the room temperature,
warm up the tubing under your
pillow
Irritation of the eyes or dry eyes Air escaping from between the
mask and the skin of your face
Check the mask for proper
positioning
Consult a physician
Redding of the skin in the area
of the mask, swelling of the
skin
Wrong mask size
Inform the attending physician
Head gear too tight
Head straps loosened
allergical reaction
Inform the attending physician