1. As soon as the device is placed properly
please connect it with the power supply cable
to an earthing contact socket.
2. Switch on the AC power switch on the back of the
device. The device welcomes you with a little text
in the display. Afterwards the display switches
over to indication of current date and time.
3. Connect the therapeutic tube
a) to the air outlet and
b) to the breathing-mask.
The device must be used with an exhalation port.
4. Connect the measuring tube
to the sensor socket (measuring tube connection).
5. Put on the breathing-mask. If the automatic
operating mode has been selected the device
will start working upon the first breath. In case
of manual operating mode you must press
the control key (ON/OFF) for starting the device.
6. Now first
TREND 300
starts with the prescribed
pressure for the certain period of time you have
selected for mask test. During this time you
should fit the mask correctly in order to eliminate
possible leakages.
14