Chapter 2: Safety warnings
Electrical safety
Risk of injury due to electric shock!
•
Do not try to open the device. Maintenance and repairs
may only be performed by personnel authorised by
HOFFRICHTER GmbH.
•
Do not touch live parts of the mains cable or power supply
unit if defective.
Ö
Replace a defective mains cable/power supply unit.
•
The device must not be used in wet rooms, as humidity
penetrating the device presents a risk of electric shock.
Risk of injury due to interrupted operation!
•
The device must always be located at least 30 cm away from
other devices or equipment such as defibrillators, diathermy
units, mobile phones, microwaves, remote controlled toys
etc. Electromagnetic fields that exceed 10 V/m may adversely
affect the operation of the ventilator.
•
During certain examinations or treatments, mutual inter-
ference between the ventilator and other medical devices
may occur. Please observe the information regarding elec-
tromagnetic compatibility and monitor the devices with
regard to error-free and proper operation.
•
The use of accessories or power supplies we have not
approved for the ventilator can increase the emission of
electromagnetic radiation, reduce interference immunity
or lead to an increased patient leakage current.
•
Only the supplied power supply unit may be used for operating the ventilator.
•
Only LAVI and a humidifier may be connected to a power strip. Additional power
strips or extension cables must not be plugged in to this power strip.
•
Portable power strips the LAVI/an external humidifier are connected to must not
be placed on the floor.
•
Do not exceed the permitted maximum load of the power strip. Refer to the user’s
manual for the maximum power consumption of the LAVI/humidifier.
•
Respiratory therapy may be contraindicated for certain pre-existing conditions.
•
The contacts for connecting the remote alarm/nurse call and the RS232 interface
must not be touched at the same time as the patient to prevent current from
being discharged via the patient.
•
Only accessories approved by HOFFRICHTER that are not connected to the power
supply may be connected to the remote alarm/nurse call and the RS232 interface.
Содержание LAVI
Страница 2: ......
Страница 18: ...18 Chapter 1 Introduction ...
Страница 26: ...26 Chapter 2 Safety warnings ...
Страница 27: ...Chapter 3 Description of device 27 Chapter 3 Description of Device ...
Страница 33: ...Chapter 3 Description of device 33 Chapter 4 Commissioning ...
Страница 51: ...Chapter 5 Device operation 51 Chapter 5 Operating the Device ...
Страница 75: ...Chapter 6 Alarms and messages 75 Chapter 7 Cleaning and Exchange of Components ...
Страница 81: ...Chapter 8 Routine checks and maintenance work 81 Chapter 8 Routine Tests and Maintenance Work ...
Страница 84: ...84 Chapter 8 Routine checks and maintenance work ...
Страница 85: ...Chapter 9 Appendix 85 Chapter 9 Annex ...
Страница 98: ...98 Chapter 9 Appendix ...
Страница 99: ...Chapter 9 Appendix 99 ...
Страница 100: ...100 Chapter 9 Appendix ...
Страница 101: ...Chapter 9 Appendix 101 ...
Страница 102: ...102 Chapter 9 Appendix ...