Chapter 4: Commissioning
43
Setup with external humidifier
Please note:
•
Use an approved humidifier according to DIN EN ISO 8185 with a humidity out-
put of > 33 mg/l.
•
Follow the manufacturer user’s manual.
•
The humidifier should be positioned below the patient and the device, so that no
water can accumulate in the patient’s lungs or in the ventilator. If water accumu-
lates in the tube circuit, we recommend using water traps.
Connect the components as follows:
FiO
2
PATIE
N
T
F
D
B
C
D
E
A
A
Tracheal cannula
B
Catheter mount tube
C
Passive exhalation valve
D
Leakage tube
E
Bacterial filter
F
Humidifier
Figure 13:
System setup, invasive ventilation with external humidifier
1. Connect all components as per Figure 13.
2. Calibrate the connected tube circuit (see page 45).
Содержание LAVI
Страница 2: ......
Страница 18: ...18 Chapter 1 Introduction ...
Страница 26: ...26 Chapter 2 Safety warnings ...
Страница 27: ...Chapter 3 Description of device 27 Chapter 3 Description of Device ...
Страница 33: ...Chapter 3 Description of device 33 Chapter 4 Commissioning ...
Страница 51: ...Chapter 5 Device operation 51 Chapter 5 Operating the Device ...
Страница 75: ...Chapter 6 Alarms and messages 75 Chapter 7 Cleaning and Exchange of Components ...
Страница 81: ...Chapter 8 Routine checks and maintenance work 81 Chapter 8 Routine Tests and Maintenance Work ...
Страница 84: ...84 Chapter 8 Routine checks and maintenance work ...
Страница 85: ...Chapter 9 Appendix 85 Chapter 9 Annex ...
Страница 98: ...98 Chapter 9 Appendix ...
Страница 99: ...Chapter 9 Appendix 99 ...
Страница 100: ...100 Chapter 9 Appendix ...
Страница 101: ...Chapter 9 Appendix 101 ...
Страница 102: ...102 Chapter 9 Appendix ...