ICE WEIGHT SCALE CALIBRATION PROCEDURE 

 

 

Turbine flowmeters, when properly used, have been found to be 
more accurate than most weight scales in determining the amount of 
product delivered.  Whenever possible use Hoffer portable Transfer 
Standard SY-14B to perform ICE system calibrations.  This 
procedure requires the use of two tankers.  One full of fresh cold 
product and a empty tanker.  Place the full tanker onto the scales.  
Hook up the piping so that the full tanker will be pumping off to the 
empty tanker.  Follow the steps below: 

1.

 

Power up the ICE unit and enter the MENU. 

2.

 

Select CALIBRATION mode. Set K-METHOD to AVERAGE.   

3.

 

Record the AVERAGE K-FACTOR value. 

4.

 

Select SYSTEM CONFIGURATION MODE. Set DELIVERY 
UNITS to mass (KG or LB). 

5.

 

Return the ICE to the OPERATING mode. 

6.

 

Pressurize and cool down the piping and flowmeter.    

7.

 

Press the 

CLEAR

 key on the ICE twice to clear the total.  

8.

 

Weigh the tanker and record its INITIAL WEIGHT.  After 
weighing the tanker start the pump.  

9.

 

To determine the TEST SAMPLE SIZE, divide the scale 
increment by the accuracy requirement.  For example, if scale 
increment is 10 pounds and the accuracy requirement is 0.5%, 
then 10 pounds / 0.005 = 2000 pounds.  It is best to use an 
accuracy requirement equal to the linearity rating of the 
flowmeter (typically 0.5%) 

10.

 

Open the discharge valve and allow the tanker on the scales to 
pump off an amount of fluid equal to the TEST SAMPLE SIZE.  
After the TEST SAMPLE SIZE is pumped, close the discharge 
valve.  Weigh the tanker again, this is the tanker FINAL 
WEIGHT.  Calculate the SCALE TOTAL by subtracting the 
FINAL WEIGHT from the INITIAL WEIGHT. 

Содержание Integrated Cryogenic Electronics HP-319

Страница 1: ...Cryogenic Electronics QUICK START GUIDE HP 319 April 2018 107 Kitty Hawk Lane P O Box 2145 Elizabeth City NC 27909 1 800 628 4584 252 331 1997 Fax 252 331 2886 www hofferflow com email info hofferflow...

Страница 2: ...HFC 9708...

Страница 3: ...be requested from HOFFER FLOW CONTROLS This document contains proprietary information which is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or...

Страница 4: ...FC will notify the purchaser in writing and will deliver the goods back to Purchaser upon i receipt of Purchaser s written instructions and ii the cost of transportation If Purchaser does not respond...

Страница 5: ...e 8 System Configuration Mode 10 THE DELIVERY PROCESS 12 Cooling the System 12 Dispensing the Product 12 Ending the Delivery 12 PRINT FUNCTIONS 13 Delivery Ticket 13 Trip Report 14 Print Configuration...

Страница 6: ...HP 319 i ICE Quick Start Guide THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 7: ...printer ICE has been designed in accordance with the cryogenic metering section of NIST HANDBOOK 44 and OIML R 81 MODEL NUMBER DESIGNATION MODEL ICE A B C D E F G SERVICE PUMP INTERLOCK PULSE SECURIT...

Страница 8: ...MODEL ICE F OPTION F S STANDARD FLAT MOUNT WITH SHOCKS SM SHOCK MOUNTED SWIVEL STICK E EX AND NEMA ENCLOSURES ND NEMA 4X ENCLOSURE FOR WALL MOUNTING SPECIAL FEATURE MODEL ICE G OPTION G SW POWER SWITC...

Страница 9: ...ket THE USER INTERFACE ICE is designed to be a user friendly component of the delivery system with an intuitive interface and easy access to the most common operations The pushbuttons allow the user t...

Страница 10: ...e may be viewed by pressing the MODE button When prompted for a password simply press MODE to advance to each of the 3 other mode screens until returning to the Delivery Mode Delivery Mode The Deliver...

Страница 11: ...ection for Delivery Total Temperature 4 1 000 0 The measured operating temperature If the temperature sensor fails or is removed the default temperature for the selected liquid is used for calculation...

Страница 12: ...he bottom portion of the screen is reserved for displaying error messages The CLEAR button clears the Delivery Total from the Delivery Mode screen The PRINT button prints a Delivery Ticket from the De...

Страница 13: ...st recent configuration changes with date time stamp for the selected number of entries Trailer Numeric Entry Enter up to an 8 digit number between 1 and 99999999 Coil Ohms Display Only Displays the t...

Страница 14: ...earization Select flowmeter linearization method as Average single K or Linearization 5 point linearization table Numeric Entry of 5 frequency and k factor points Enter the 5 point k factor table Freq...

Страница 15: ...nly Displays the maximum flowrate in gallons per minute for the selected meter size Min Delivery lb Display Only Displays the minimum delivery in pounds for the selected meter size ICE SN Numeric Entr...

Страница 16: ...lid entry is between 20 degrees Kelvin and the max temperature in the compensation range for a selected fluid Default Press Numeric Entry Enter the default value for the fluid temperature to be used i...

Страница 17: ...s liters ft3 ft3X100 m3 pounds kilograms Select the units of measure for Delivery Total Flowrate will be displayed in the same units NOTE If Compensation Method is set to NONE only gallons and liters...

Страница 18: ...layed but not recorded in the Trip Log After the first minute of the delivery any error occurring more than one minute is recorded in the Trip Log See section Error Messages for a comprehensive list o...

Страница 19: ...nt functions Message Description Printer out of paper There is no paper detected in the printer Printer offline no response from printer The printer power is off or the serial cable is not connected b...

Страница 20: ...It can be thought of as series of delivery tickets with the addition of any errors that were logged during each delivery From the Maintenance Mode screen press the Print Trip Report button on the disp...

Страница 21: ...nce Mode screen press the Audit Trail button on the display to bring up a screen to select the number of most recent entries to print A password is not required to print the Audit Trail Sample Audit T...

Страница 22: ...in the Turbine Calibration Mode This message may appear if the turbine is being spun by gas LOW FLOW RATE The flow rate has dropped below the limit defined by the Min Flow field for the selected mete...

Страница 23: ...ROR The number of missing or double pulses detected has exceeded the acceptable percentage of delivery total PULSE ERROR EXCEEDS LIMIT This message is displayed on the Delivery Ticket only and indicat...

Страница 24: ...he shock mounts is 105 lbs in Hoffer recommends powering ICE directly from the truck battery The DC power line may be wired through an external switch if desired or ICE may be ordered with an optional...

Страница 25: ...224cm x 16 61cm x 12 38cm Weight 4 2 lbs 1 91 kg Approvals Design Approvals Standards and Regulatory Compliance CE Emissions EN55022 FCC Part 15 Immunity EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4...

Страница 26: ...V lithium coin cell for real time clock 2 4 years typical battery life Flow Meter Input Frequency range 0 2 to 5000 Hz Amplitude 15mV RMS to 50V RMS Impedance 10 k Linearization Single K Factor or 5 p...

Страница 27: ...Password protection to prevent unauthorized configuration changes Diagnostic Features Failure detection for RTD analog and flow meter inputs Multiple error messages Tracking of pump operational hours...

Страница 28: ...HP 319 22 ICE Quick Start Guide THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 29: ...HP 319 23 ICE Quick Start Guide APPENDIX A DRAWINGS...

Страница 30: ...HP 319 24 ICE Quick Start Guide...

Страница 31: ...HP 319 25 ICE Quick Start Guide...

Страница 32: ...HP 319 26 ICE Quick Start Guide...

Страница 33: ...assis WHT TB2 1 Signal SCA XLT2 S Signal Input BLK TB2 2 Signal WHT TB4 1 Signal SCA XLT2 S Secondary Signal Input BLK TB4 2 Signal WHT TB2 5 RTD Sense SCA XLT2 TA Temperature Probe BLK TB2 4 RTD Sens...

Страница 34: ...HP 319 28 ICE Quick Start Guide...

Страница 35: ...HP 319 29 ICE Quick Start Guide...

Страница 36: ...HP 319 30 ICE Quick Start Guide...

Страница 37: ...HP 319 31 ICE Quick Start Guide...

Страница 38: ...HP 319 32 ICE Quick Start Guide...

Страница 39: ...HP 319 33 ICE Quick Start Guide...

Страница 40: ...HP 319 34 ICE Quick Start Guide THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 41: ...HP 319 35 ICE Quick Start Guide APPENDIX B ICE WEIGHT SCALE CALIBRATION PROCEDURE...

Страница 42: ...N MODE Set DELIVERY UNITS to mass KG or LB 5 Return the ICE to the OPERATING mode 6 Pressurize and cool down the piping and flowmeter 7 Press the CLEAR key on the ICE twice to clear the total 8 Weigh...

Страница 43: ...R AVERAGE_K_FACTORNEW DELIVERY_TOTAL ________________ SCALE_TOTAL x AVERAGE_K_FACTOROLD 12 Enter the ICE CALIBRATION mode and enter the new AVERAGE K FACTOR 13 Repeat Steps 6 through 9 to verify the n...

Отзывы: