Battery Information 

The logger is installed with a 3V CR2032 battery (HRB-TEMP). 
Expected battery life varies based on the ambient temperature 
where the logger is deployed, the logging interval, the rate of 
state changes, the frequency of offloading to the computer, 
number of channels that are active, and battery performance. A 
new battery typically lasts 1 year with a logging interval greater 
than 1 minute. Deployments in extremely cold or hot 
temperatures or a logging interval faster than 1 minute may 
reduce battery life. Estimates are not guaranteed due to 
uncertainties in initial battery conditions and operating 
environment. 

The logger can also be powered by the USB cable when the 
remaining battery voltage is too low for it to continue logging. 
Connect the logger to the computer, click the Readout button 
on the toolbar, and save the data as prompted. Replace the 
battery before launching the logger again. 

To replace the battery: 

1.

 

Open the battery door on the top of the logger. 

2.

 

Use the tab on the built-in battery removal tool inside the 
compartment to remove the battery. 

3.

 

Place the circular battery removal tool around the negative 
side of the new battery with the tab up. Insert the new 
battery with the positive side facing out. The LCD should 
display “HOBO” briefly after correctly installing the battery. 

 

 

WARNING:

 Do not cut open, incinerate, heat above 85°C 

(185°F), or recharge the lithium battery. The battery may 
explode if the logger is exposed to extreme heat or conditions 
that could damage or destroy the battery case. Do not dispose 
of the logger or battery in fire. Do not expose the contents of 
the battery to water. Dispose of the battery according to local 
regulations for lithium batteries. 

HOBOware provides the option of recording the current battery 
voltage at each logging interval, which is disabled by default. 
Recording battery life at each logging interval takes up memory 
and therefore reduces logging duration. It is recommended you 
only record battery voltage for diagnostic purposes. 

Battery door

CR2032 battery; positive  

side facing out 

Tab on battery-removal tool

Mounting magnets 

Mounting magnets

Содержание UX90-005

Страница 1: ...Memory Modes Wrap when full or stop when full Start Modes Immediate push button date time or next interval Stop Modes When memory full push button or date time Time Accuracy 1 minute per month at 25 C 77 F see Plot A Battery Life 1 year typical use Battery Type One 3V CR2032 lithium battery Memory UX90 005 006 128 KB 84 650 measurements maximum UX90 005M 006M 512 KB 346 795 measurements maximum Do...

Страница 2: ...ecording Internal Logger Events or to turn the LCD screen on if the option to turn off the LCD has been enabled see Setting up the Logger Battery door Open the battery door not visible in the diagram on the top of the logger to access the logger battery see Battery Information Calibration Button Press this button to calibrate the logger for the light you will be monitoring See Calibrating the Ligh...

Страница 3: ... has been on since logging began ranging from seconds to days This example indicates the room has been occupied or the light has been on for a total of 5 minutes and 38 seconds The logger must be launched with the LCD set to show Time for this symbol to display Time display when logger is stopped This indicates the logger has been configured to start logging on a particular date time The display w...

Страница 4: ...eading out the logger 6 Set the units to display on the LCD screen Select either Time or 7 If the logger is configured to record runtime choose a logging interval from 1 second to a maximum of 18 hours 12 minutes and 15 seconds 8 Choose when to start logging Now Logging begins immediately At Interval Logging will begin at the next even interval available when logging runtime only On Date Time Logg...

Страница 5: ... Recording battery voltage reduces storage capacity and is generally not used except for troubleshooting Calibrating the Light Sensor Each time you place the logger in a new lighted environment you should calibrate it to the light level that you will be monitoring This ensures the logger is accurately determining when the light changes between ON and OFF states There are two calibration methods av...

Страница 6: ...p recording data from HOBOware or by pushing the Start Stop button Internal Calibration The logger was calibrated via auto calibration button calibration Host Calibration The logger was calibrated via HOBOware Calibration Failure Calibrating the logger has failed Button Up Button Down The Start Stop button was pressed for 1 second Safe Shutdown The battery level dropped below 2 5 V the logger perf...

Страница 7: ...mputer click the Readout button on the toolbar and save the data as prompted Replace the battery before launching the logger again To replace the battery 1 Open the battery door on the top of the logger 2 Use the tab on the built in battery removal tool inside the compartment to remove the battery 3 Place the circular battery removal tool around the negative side of the new battery with the tab up...

Отзывы: