2
-
4.
Remove the 3 desiccant packs and inspect them. If they are not bright blue, dry them by leaving
them in a warm (
≤
70
o
C (158
o
F), dry location until the bright blue color is restored. If a desiccant
pack remains pink and will not turn blue, replace it with a new desiccant pack (Onset Part #:
DESICCANT1).
5.
Using a utility knife, cut the tie holding the sensor to the board.
6.
Disconnect the cable from the jack on the board (pull straight out).
7.
Pull the cable out through the case.
8.
Discard cable, dome nut, and grommet.
Connecting the New Cable/Sensor
1.
Slide the new cable through the case, connector end first.
2.
Apply a small amount of grease from included grease packet to the O-ring on the grommet, just
enough to moisten the ring.
3.
Slide the dome nut/grommet assembly up towards the case and hand-tighten the grommet. Do not
tighten the dome nut until assembly is complete to allow for sliding of the cable.