background image

HOBO U20 Water Level Logger Manual 

 

 

1-800-LOGGERS (564-4377) • 508-759-9500 

www.onsetcomp.com/support/contact 

© 2008–2017 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOware are trademarks or 
registered  trademarks  of  Onset  Computer  Corporation.  All other trademarks are the property of their respective 
companies. 

This product has been manufactured by Onset Computer Corporation and in compliance with Onset’s ISO 9001:2008 
Quality Management System. 

Patent #: 6,826,664 

12315-H MAN-U20 

 

The black acetal cap is provided to help protect the 
communications window. Acetal is resistant to most solvents, 
fuels, and lubricants. 

The polycarbonate communications window is sealed as an 
additional barrier to water and dirt entering the logger housing. 

Compensating for Drift 

All pressure sensors drift over time. The drift for the pressure 
sensor and electronics in the HOBO Water Level logger is less 
than 0.3% FS (worst case) per year. In most applications, drift is 
not a significant source of error, because the offset created by 
any drift is zeroed out when you take a manual reference level 
measurement and use the logger software to automatically 
calculate the level readings relative to the reference 
measurement. In effect, you are re-zeroing the sensor each 
time you apply a reference reading to the data file. 

Pressure sensor drift matters only when absolute pressure 
values are needed, or if there are no recent reference level or 
depth measurements available. For example, if the logger is 
deployed for one year and no new reference level readings are 
taken during the deployment, it is possible that the sensor 
could have drifted as much as 0.3% FS by the end of the 
deployment.  

It is possible to determine the actual amount of drift during a 
deployment if a reference level is taken at the beginning and 
the end of a long-term deployment. The results of applying the 
two different reference levels (once at the beginning of the 
data file, and again at the end of the data file) can be 
compared. Any difference between the files indicates the 
amount of sensor drift (assuming accurate reference levels). 

Verifying Accuracy 

You can check the 

differential accuracy

 of your loggers for 

water level measurements by deploying the loggers at two 
depths and comparing the difference in level readings. When 
verifying the accuracy this way, be sure to allow the loggers’ 
temperature to stabilize at each depth. Use the logger software 
to convert the readings from pressure to level. The level 
readings should be taken close enough together that the 
barometric pressure does not change.  

You can check the 

absolute pressure accuracy

 of your HOBO 

Water Level Logger by comparing its ambient pressure readings 
to a second HOBO logger. Their readings should be within each 
other’s specified accuracy. Alternatively, you can check the 
pressure reading against an accurate local barometer. If you 
use a non-local source of barometric information, such as the 
NOAA website, adjust for altitude.  

Recalibration 

If you would like to have your logger’s absolute accuracy 
verified against a NIST standard, or to have your logger 
recalibrated, contact Onset or your place of purchase for 
pricing and return arrangements. 

The Battery 

The battery in the HOBO Water Level Logger is a 3.6 Volt 
lithium battery.  

Battery Life 

The battery life of the logger should be about five years or 
more. Actual battery life is a function of the number of 
deployments, logging interval, and operation/storage 
temperature of the logger. Frequent deployments with logging 
intervals of less than one minute, and continuous 
storage/operation at temperatures above 35°C will result in 
significantly lower battery life. For example, continuous logging 
at a one-second logging interval will result in a battery life of 
approximately one month. 

To obtain a five-year battery life, a logging interval of one 
minute or greater should be used and the logger should be 
operated and stored at temperatures between 0° and 25°C (32° 
and 77°F).  

Voltage 

The logger can report and log its battery voltage. If the battery 
falls below 3.1 V, the logger will record a “bad battery” event in 
the datafile. If the datafile contains “bad battery” events, or if 
logged battery voltage repeatedly falls below 3.3 V, the battery 
is failing and the logger should be returned to Onset for battery 
replacement.  

Replacing the Battery 

To have your logger’s battery replaced, contact Onset or your 
place of purchase for return arrangements. Do not attempt to 
replace the battery yourself. Severe damage to the logger will 
result if the case is opened without special tools, and the 
warranty will be voided. 

 

WARNING:

 Do not cut open, incinerate, heat above 

100°C (212°F), or recharge the lithium battery. The battery 
may explode if the logger is exposed to extreme heat or 
conditions that could damage or destroy the battery case. 
Do not dispose of the logger or battery in fire. Do not 
expose the contents of the battery to water. Dispose of the 
battery according to local regulations for lithium batteries. 

Содержание U20

Страница 1: ...nt accuracy also depends on temperature response time Pressure Absolute and Water Level Measurements U20 001 02 and U20 001 02 Ti Operation Range 0 to 400 kPa 0 to 58 psia approximately 0 to 30 6 m 0...

Страница 2: ...olt lithium factory replaceable Battery Life Typical Use 5 years with 1 minute or greater logging interval Memory Non volatile 64K bytes memory approx 21 700 pressure and temperature samples Weight St...

Страница 3: ...sistant for easy and accurate barometric compensation LEDs A light LED in the communications window of the logger confirms logger operation The following table explains when the logger blinks during l...

Страница 4: ...ny environments The logger uses an absolute pressure sensor so no vent tube is required The small size of the logger is convenient for use in small wells and allows the logger to be mounted and or hid...

Страница 5: ...point such as referencing groundwater measurements to the top of the well record the water level as a negative number If the water level surface is above the reference point such as height above sea l...

Страница 6: ...drift or change in the suspension cable length To quantify measurement error which is ideally zero compare the calculated water level at the end of the plot with the water level measured just before y...

Страница 7: ...th Use the logger software to convert the readings from pressure to level The level readings should be taken close enough together that the barometric pressure does not change You can check the absolu...

Отзывы: