background image

 

HOBO MX CO

2

 Data Logger (MX1102A) Manual 

   

 

 

Auto calibration.

 Use this option if you want the logger to 

automatically calibrate within the first 24 hours after 
logging begins and then every eight days thereafter. The 
logger will be calibrated based on the average of the three 
CO

2

 measurements that follow the lowest CO

2

 value 

identified during the 24-hour or 8-day time period as 
applicable. 

Important:

 Accurate auto-calibration requires the building 

or location where the logger is deployed to be empty at 
least once during the eight-day period (for example, an 
empty office building during the weekend or overnight will 
typically have background CO

2

 levels of 400 to 450 ppm). 

If the logger is deployed in an area where the CO

2

 level 

does not go down to 400 ppm during the eight-day time 
period, then manual calibration should be performed 
regularly instead or inaccurate CO

2

 readings will be 

reported. If you plan on using auto calibration but the 
building will be occupied during the first day after logging 
begins, then you can use the manual calibration option as 
well. You can manually calibrate the logger immediately 
after logging begins and use auto calibration thereafter. 

Note:

 Every time the logger is started, auto calibration will 

occur after 24 hours and then again after eight days unless 
a manual calibration is performed first. 

 

Altitude compensation.

 The CO

2

 sensor must compensate 

for locations above or below 305 meters (1,000 feet) to 
provide an accurate reading. You must enter the altitude 
above or below sea level in either meters or feet when 
configuring the logger if it will be deployed at a location 
above or below sea level. In normal use, the CO

2

 

measurement will vary by approximately 0.135% of the 
reading for each mbar change in barometric pressure (the 
sensor is calibrated at 1,013 mbar). Use altitude 
compensation when deploying the logger for the best CO

2

 

accuracy possible. 

To access calibration and altitude compensation settings in 
the app: 

1.

 

Tap Devices and then tap the logger in the app to connect 
to it. 

2.

 

Tap . 

3.

 

Tap the CO

2

 sensor. 

4.

 

Select auto calibration, manual calibration, or both. 

5.

 

Select “Altitude Compensation” and enter the altitude 
above or below sea level where the logger will be deployed 
in either meters or feet. 

6.

 

Tap Save. 

7.

 

Tap . 

To access calibration and altitude compensation settings in 
HOBOware: 

1.

 

Connect the logger to the computer with the USB cable. 

2.

 

From the Device menu in HOBOware, select Launch. 

3.

 

Click the CO

2

 Settings button. 

4.

 

Select auto calibration, manual calibration, or both. 

5.

 

Select “Use Carbon Dioxide sensor altitude compensation” 
and enter the altitude above or below sea level where the 
logger will be deployed in either meters or feet. 

6.

 

Click OK. 

7.

 

Click Start in the Launch Logger window to load settings to 
the logger. 

Note: 

If both auto calibration and manual calibration are 

selected, the logger will automatically calibrate within 24 hours 
after logging begins unless a manual calibration occurs during 
that time period. In addition, when both calibration settings are 
selected, the eight-day calibration cycle will be reset any time a 
manual calibration is performed. 

To manually calibrate the CO

2

 sensor: 

Important:

 If you do not follow these manual calibration 

instructions as described, the sensor readings will be incorrect 
and you will need to manually calibrate the logger again. 

1.

 

Take the logger outside in fresh air on a dry day where the 
carbon dioxide level is 400 ppm. You can also use an indoor 
location for manual calibration if it is unoccupied and is not 
exposed to a ventilation system. 

2.

 

Press the Calibrate button on the logger for 5 seconds until 
it beeps. The logger will then calibrate for 5 minutes. The 
CO

2

 and Calibrate symbols on the LCD will flash while the 

calibration is underway. A time- and date-stamped manual 
calibration event is logged in the data at the end of the 5-
minute calibration sequence. 

3.

 

Once the Calibration process is complete, return the logger 
to its deployment location. Repeat this process at least once 
every eight days for best accuracy. 

Setting up Alarms 

You can set an alarm to trip on the logger when a sensor 
reading rises above or falls below a specified value. This can 
alert you to problems so you can take corrective action. To set 
up a sensor alarm: 

1.

 

In the app: Connect to the logger and tap 

In HOBOware: From the device menu, select Launch. Click 
the Alarms button in the Launch Logger window. 

2.

 

Select the sensor that you want to set up with an alarm 
condition. 

3.

 

Enable the High Alarm if you want an alarm to trip when the 
sensor reading rises above the high alarm value. Drag the 
slider to the reading that will trip the alarm or type a 
specific reading. 

4.

 

Enable the Low Alarm if you want an alarm to trip when the 
sensor reading falls below the low alarm value. Drag the 
slider to the reading that will trip the alarm or type a 
specific reading. 

5.

 

Set the duration before an alarm is tripped and select one 
of the following: 

 

Cumulative. The alarm will trip once the sensor reading is 
out of the acceptable range for the selected duration any 
time during logging. For example, if the high alarm is set 
to 85°F and the duration is set to 30 minutes, then the 
alarm will trip once the sensor readings have been above 

www.

.com

[email protected]

1.800.561.8187

Содержание MX1102A

Страница 1: ...f 4 5 including hysteresis at 25 C 77 F below 20 and above 80 6 typical Resolution 0 01 Drift 1 per year typical CO2 Sensor Range 0 to 5 000 ppm Accuracy 50 ppm 5 of reading at 25 C 77 F less than 90 RH non condensing and 1 013 mbar Warm up Time 15 seconds Calibration Auto or manual to 400 ppm Non linearity 1 of FS Pressure Dependence 0 13 of reading per mm Hg corrected via user input for elevatio...

Страница 2: ...e reduced by remaining connected excessive readouts audible alarms and paging Visual audible alarms and other events can have a marginal impact on battery life Memory 128 KB 84 650 measurements maximum Download Type USB 2 0 interface or via Bluetooth Smart Full Memory Download Time 20 seconds via USB approximately 60 seconds via Bluetooth Smart may take longer the further the device is from the lo...

Страница 3: ...tooth The more bars there are the stronger the Bluetooth signal The logger is currently communicating with HOBOware via USB cable The logger is currently being powered by USB cable The battery indicator shows the approximate battery power remaining The logger has been configured to stop logging when memory fills The memory bar indicates the approximate space remaining in the logger to record data ...

Страница 4: ...en in the software When the LCD is turned off for logging you can still temporarily view the LCD screen by pushing the Start Stop or Clear Next button The LCD will then remain on for 10 minutes When the logger is connected to the computer with the USB cable the LCD screen refreshes every second regardless of logging interval When the logger has stopped logging the LCD screen will remain on with ST...

Страница 5: ...ice attempts to connect to it To reset a password simultaneously press both the Start Stop button and the Clear Next button on the top of the logger for 3 seconds or tap and tap Reset Mark the logger as a favorite You can then filter the list of devices to only show loggers marked as favorites Update the firmware on the logger A logger readout will be completed automatically at the beginning of th...

Страница 6: ... rate See Statistics Logging for more information Show LCD Enable or disable to control whether the LCD remains illuminated while the logger is logging If you disable the LCD the logger will not show the current reading status or other information while the logger is logging You will however be able to temporarily turn the LCD screen on by pressing the Start Stop button on the logger for 1 second ...

Страница 7: ...ect Use Carbon Dioxide sensor altitude compensation and enter the altitude above or below sea level where the logger will be deployed in either meters or feet 6 Click OK 7 Click Start in the Launch Logger window to load settings to the logger Note If both auto calibration and manual calibration are selected the logger will automatically calibrate within 24 hours after logging begins unless a manua...

Страница 8: ...lution specifications When you read out the logger alarm events can be displayed on the plot or in the data file See Logger Events Once cleared an audible alarm will start beeping again if the sensor values go out of the normal range Even if an audible alarm is cleared a visual alarm may remain on the logger LCD and in the app if applicable depending on the settings selected for maintaining visual...

Страница 9: ... following statistics can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values and The standard deviation from the average for all sampled values For example a logger is configured with the temperature and CO2 sensors enabled and the logging interval set to 5 minutes The current reading and all four statistics are enabl...

Страница 10: ...am are not automatically available in the other To open app files in HOBOware share the HOBO file via email and open it on the computer with HOBOware Files in HOBOware cannot be viewed in the app You can however export data in HOBOware to a text or Excel file that you can open on your mobile device See the app user s guide and HOBOware Help for details on sharing or exporting data Logger Events Th...

Страница 11: ...es can impact battery life Estimates are not guaranteed due to uncertainties in initial battery conditions and operating environment The logger can also be powered by the USB cable if faster logging intervals are desired or when the remaining battery voltage is too low for it to continue logging Connect the logger to the computer click the Readout button on the toolbar in HOBOware and save the dat...

Отзывы: