We can use same operate enter the resolution setting.
2.5.2
CALIBRATION SETTING
After setting the capacity, we must set the calibration point for keeping best liner
ability.
1.One point calibration means we set one point calibration weight.
We advise you
choose calibration point ≥20%*Capacity, it can be reduced error, you’d better
choose the full range as one point calibration
2.MAX: Setting the full capacity weight. (Keep first number,6200g will set 6000g)
3.HALF: Setting 50% weight value. (Example: capacity 6200g, we set 50%:3000g)
4.MIN: Setting small point 30% capacity, the better way is one weight can meet
require, for example 6200g, we can set 1kg or 2kg.