17
) Q1E-EP1370
WARNING
The sterilization method and operating conditions are as follows.
Sterilization Method
Condition
Plasma Sterilization:
STERRAD® 50, 100S or 200 (*)
Short Cycle
Plasma Sterilization:
Sterrad® NX or 100NX (*)
Standard cycle
ETO Sterilization
Gas Type: 10% EO/ 90% HCFC
Temperature: 50-55°C
Exposure Time: More than 120 minutes
Pressurization: 162-200kPa
Depressurization: 13-8kPa
Relative humidity: 40-90%
Aeration is minimum 12 hours
*STERRAD® systems are manufactured by "Johnson & Johnson"
1)
Before performing sterilization, check that the operation data of
sterilizer are in conjunction with min. and max. data applicable for
the probe and elasto coupler.
2)
Do not sterilize the probe and elasto coupler by Steam Autoclaving.
If you autoclave it, it suffers serious damage and will be not
functional
The packaging procedure is as follows.
1)
Put the probe into TYVEK pouch.
TYVEK Pouch
Fig. 6 Packaging
in the pouch