- 23 -
Q1E-EP1199
probe.
2)
Immerse the magnetic sensor attachment and the spacer for EZU-RV2S
into the disinfectant (see Fig. 17). Set a clock to insure the
recommended contact time which is 5 minutes.
3)
Rinse the magnetic sensor attachment and the spacer for EZU-RV2S with
deionized water for 1 minute. (alternatively: immerse them in a tray
filled with deionized water (see Fig. 17) for 5 min.)
4)
Visually check the outer surface of the magnetic sensor attachment
and the spacer for EZU-RV2S for leavings of the disinfectant. If
necessary, repeat the rinsing.
Water
Detergent
Disinfectant
Fig. 17 Immersion of the probe, the magnetic sensor attachment,
the spacer for EZU-RV2S and the elasto Coupler