- 17 -
Q1E-EP1199
5. Reprocessing
Procedure
The probe, magnetic sensor attachment and spacer for EZU-RV2S must be
reprocessed after each use. Refer to the reprocessing instruction in
this chapter.
Please refer to the instruction manual of EZU-TEATC2 about the method
of reprocessing EZU-TEATC2 and EZU-TECPL1.
Please refer to the instruction manual of the needle guide bracket
EZU-PA7L3 about the method of reprocessing it.
WARNINGS
‐
The probe is delivered unsterile. Prior to the first
use, reprocess the probe.
‐
Temperature should not exceed 60°C during
reprocessing.
‐
Probe connector is not water resistant.
Limitations on
reprocessing
The probe is not completely submersible. The
immersible part is shown in Fig.1. The un-immersible
part should be disinfected by wipe disinfection.
Transportation
before using
The probe should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful not
to damage the sterile pouch or container during
transportation.