7.1
Probe
Item EUP-L53S
EUP-L52
EUP-L53
EUP-L53L
Center frequency
7.5MHz 5.5MHz 7.5MHz 7.5MHz
Type
Linear array probe
Technology
High density linear array probe
Dimensions
See Fig. 11
See Fig. 9 See Fig. 10
Weight
Approx. 1.5kg (incl. Cable and connector)
Probe materials
Bio-compatible allergy free components
Acoustic output
According to IEC60601-2-37 (See Main Unit manual.)
Applicable systems
Depending on production and upgrade status. For
detailed information contact a service support.
Classification
MDD classification IIa.
Cleaning
Applicable disinfectants are listed in the suppliers
list
Sterilization
ETO gas sterilization / Plasma sterilization
Operating
conditions
Ambient temperature : 25 - 35°C
Contact surface temperature
(temperature of examinee) : max. 42°C
Relative humidity : 30 - 85%
Storage conditions
Temperature : -10 - +55°C
Relative humidity : 10 – 95%
(subject to no condensation)