MN1-5821 Rev. 4
Figure 15 Release of the probe from the biopsy attachment during puncture
Warning
Use only sterile ultrasound gel.
The use of non-sterile ultrasound gel could result in infection to the patient.
Use only the biopsy attachment specified in this manual.
A puncture with the other attachment can result in puncture at unintended parts.
Sterilize the probe after use, if the probe cover was broken during use.
Sterilize the probe according to the separate instruction manual “Cleaning, Disinfection and
Sterilization“, otherwise there is a risk of infection to the patient.
Do not reuse the probe cover that was used.
Reuse of the probe cover can result in infection to the patient.
Caution
Do not operate the probe with excessive force.
Use with excessive force could result in injury to the patient.
Scan for minimum time necessary at the lowest possible acoustic output.
Acoustic output may affect the patient’s internal tissues.
For details about the acoustic output, please refer to the documentation supplied with the ultrasound
diagnostic instrument.
Do not touch the connector terminal pin of the probe.
Electrostatic discharge may result in malfunction of the probe.
Do not touch the probe connector of the ultrasound diagnostic instrument and the patient at the same
time. It can cause electric shock to the patient.
①
Push the lever of
the Needle Insert.
①
①
②
③
②
③
Needle Insert
Needle
Slide the Needle Insert
upward.
Remove the Needle Insert.
Содержание C25P