98
Message Reference
(4) Selecting the channel
Select and query channel
Syntax
Command
[ROUTe]: CLOSe <
Slot and channel No.
>
Query
[:ROUTe]:CLOSe?
Response
<Slot and channel No.>
<Slot and channel No.> = Slot No. × 100 + CH No.
A 3-digit number (or 4-digit number) is used to specify the channel.
Example: 208 (or 0208) when specifying channel 8 for
SLOT 2
.
1112 when specifying channel 12 for
SLOT 11
.
Description
Closes the specified slot and channel. The channel that was closed previously is
automatically opened. You can use the
*
OPC?
query to check whether close operation has
completed.
Example
:CLOS 0122
Channel 22 for
SLOT 1
is closed.
:CLOS?
905
Channel 5 for
SLOT 9
is closed.
Note
• Specifying a slot without a module will result in an execution error.
• An execution error occurs during scan operation (command).
Open all relays
Syntax
Command
[:ROUTe]:OPEN
Description
Opens all channels. You can use the
*
OPC?
query to check whether close operation has
completed.
Example
Command
:OPEN
Note
Executing this command while scan operation is in progress will cause scan operation to
stop and all relays to open.
Set and query channel delay time
Syntax
Command
:SYSTem:MODule:DELay <
Slot No.
>,<
Time/MIN/MAX/DEF
>
Query
:SYSTem:MODule:DELay?
Slot No.
>
Response
<Time>
<Slot No.> = 1 to 3 (NR1) (SW1001)
1 to 12 (NR1) (SW1002)
<Time> = 0 to 9.999 (NR2) [s] (MIN:0, MAX: 9.999, DEF: 0)
Description
Sets the additional delay time after relay operation (standby for the specified settling time
occurs).
After the delay time elapses, channel closing is completed.
The channel delay time can be specified for each slot.
Example
:SYSTem:MODule:DELay 1,0.01
Set the channel delay time for
SLOT 1
to 0.01 s.
:SYSTem:MODule:DELay? 3
0.5
The channel delay time for
SLOT 3
has been set to 0.5 s.
Note
• Specifying a slot without a module will result in an execution error.
• An execution error occurs during scan operation (command).
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...