41
Fig. 9.1.1
Fig. 9.1.1 shows the DC OUT (OPT ION) “G”/”+” terminals on the
external input/outp ut terminal b lock on the rear of the unit.
A DC measuring instrument like a HI OKI chart recorder can be
connected to these terminals. For permanent data record ing, a HIOKI
chart recorder is recommended for better technical follow-up, includ ing
a supply of a variety of recording charts and technical service.
3) Output voltage
These tables show the relations between disp layed value of super
megohmmeter and output voltage of RP-8000.
Target voltage 5V
Disp layed value
0.025
0.125
0.25
Output voltage
1V
5V
10V
Target voltage 10V
Disp layed value
0.05
0.25
0.5
Output voltage
1V
5V
10V
Target voltage 15V
Disp layed value
0.075
0.375
0.75
Output voltage
1V
5V
10V
Target voltage 25V
Disp layed value
0.125
0.625
1.25
Output voltage
1V
5V
10V
Target voltage 50V
Disp layed value
0.25
1.25
2.5
Output voltage
1V
5V
10V
Target voltage 100V
Disp layed value
0.5
2.5
5
Output voltage
1V
5V
10V