18
4.4 Measuring Condition Setting Display
The LCD display showing the measuring condition setting.
Fig. 4.4
⑩
Status and Mode Notice:
When the MEASURE/DISCHARGE selector
switch is set to the DISCHARGE position, press the SELECT switch to
enter the setting mode, and the selected mode is shown in this box.
Each time the SELECT switch is pressed, the mode is changed in
the order;
C. SET
(voltage charging time setting mode)
→
M. SET
(measuring time setting mode)
→
COMP (measured value comparison
level setting mode)
→
BUZZ (buzzer sound level setting mode)
→
(blank – measuring display)
.
When the RANGE selector is set to the CAL position, CAL1
is shown.
With the CAL1 status shown, the press of the SELECT switch changes
the status to CAL2 .
⑪
Highlighted Comparison and Judgment Value: When the COMP (comparison
and judging level setting mode) is selected, the value is shown high-
lighted, and the it can be changed with the UP and DOWN switches.
⑫
Comparison and Judgment Value Mark: This heart mark is shown when the
comparison and judgment function is set to on, showing the position
of the comparison and judgment level.
⑬
Highlighted Judgment value: This shows the judgment level which can be
changed with the UP/DOWN switches.
⑭
LOCK (Interlocking) Notice: This shows that preparation for the measure-
ment is not ready, yet. This notice is shown when the interlocking
function is used to show that the HV-EN connector is free, and/or the
Rx
/
+
measuring terminals are free.
When the LOCK notice is shown, the MEASURE and CHARGE switches
become inoperative.
LOCK
100V
⑭
⑬
⑫
⑪
⑩
3
M.S ET
s
MΩ
10
20
10
5
2
1
.5
000