 Conditions of guaranteed accuracy

• 

Temperature & humidity: 23 ˚C ±5 ˚C,  80% rh or less (non-condensating)

• 

From 0°C to 18°C and from 28°C to 40°C, add (temperature coefficient 

±[1/10 measurement accuracy] / °C).

• 

Guaranteed Accuracy Period: 1 year

• 

RM3545 only: Warmup time of 60 min. or greater (If less than 60 min., 

double figures in the accuracy table to obtain the measurement accuracy.)

• 

RM3545 only:

 

self-calibration AUTO

*When using manual self-calibration, temperature fluctuations after performing calibration must be within ±2°C, and the calibration interval must be within 30 min.

 

Measurement accuracy

RM3544

RM3545

t

0

 :  Reference temperature. [ºC]

t :  Ambient temperature. [ºC]

Δt : Temperature. measurement 

accuracy

α

t

0

 : Temperature. coefficient at t

0

 is 

[1/ºC]

-

α

t

0

Δt

1

+

α

t

0

×

(

t

+

Δt

-

t

0

)

×

100 [%]

* During temperature correction, the value calculated below is added to 

the rdg. error for resistance measurement accuracy:

Accuracy = ±(% rdg. + % f.s.)

  LP OFF

 

• f.s. = calculated 1,000,000 dgt., where 0.001% f.s. = 10 dgt. 

 

• 

For 100 MΩ and greater ranges with 100 MΩ range high-precision mode off, calculate as f.s. = 10,000 dgt. and 0.01% f.s. = 1 dgt.

Range

100MΩ 

range high-

precision 

mode

Max.  

measurement  

display 

*1

Reso-

lution

Accuracy   %rdg. + %f.s. 

*2

Measurement 

current 

*3

Additional 

accuracy 

without 

0ADJ

%f.s. 

*2

Max open-

terminal 

voltage

FAST

MED

SLOW1

SLOW2

Switching

10mΩ

  12.000 00  mΩ

10 nΩ

0.060+0.050

0.060+0.020

0.060+0.020

1A

0.020

5.5V 

*4

(0.060+0.015)

(0.060+0.002)

(0.060+0.001)

( – )

100mΩ

  120.000 0  mΩ

100 nΩ

0.060+0.010

0.060+0.010

0.060+0.010

High

1A

0.002

(0.060+0.003)

(0.060+0.001)

(0.060+0.001)

( – )

0.014+0.050

0.014+0.020

0.014+0.020

Low

100mA

0.020

(0.014+0.015)

(0.014+0.002)

(0.014+0.001)

( – )

1000mΩ

  1200.000  mΩ

1 μΩ

0.012+0.010

0.012+0.008

High

100mA

0.002

(0.012+0.003)

(0.012+0.001)

( – )

0.008+0.050

0.008+0.020

Low

10mA

0.020

(0.008+0.015)

(0.008+0.002)

( – )

10Ω

  12.000 00   Ω

10 μΩ

0.008+0.010

0.008+0.008

High

10mA

0.002

(0.008+0.003)

(0.008+0.001)

( – )

0.008+0.050

0.008+0.020

Low

1mA

0.020

(0.008+0.015)

(0.008+0.002)

( – )

100Ω

  120.000 0    Ω

100 μΩ

0.007+0.005

0.007+0.002

0.007+0.001

High

10mA

(0.007+0.005) (0.007+0.001)

(0.007+0.001)

( – )

0.008+0.010

0.008+0.010

Low

1mA

0.002

(0.008+0.003)

(0.008+0.001)

( – )

1000Ω

  1200.000   Ω

1 mΩ

0.007+0.005

0.006+0.002

0.006+0.001

1mA

(0.007+0.005) (0.006+0.001)

(0.006+0.001)

( – )

10kΩ

  12.000 00  kΩ

10 mΩ

0.008+0.005

0.007+0.002

0.007+0.001

1mA

20V

100kΩ

  120.000 0  kΩ 100 mΩ 0.008+0.005

0.007+0.002

0.007+0.001

100μA

1000kΩ

  1200.000   kΩ

1 Ω

0.015+0.005

0.008+0.002

0.008+0.001

10μA

10MΩ

  12.000 00  MΩ

10 Ω

0.030+0.005

0.030+0.002

0.030+0.001

1μA

100MΩ

ON

  120.000 0  MΩ

100 Ω

0.200+0.005

0.200+0.002

0.200+0.001

100nA

OFF

  120.00   MΩ

10 kΩ

10.00MΩ or less :  0.50+0.02

10.01MΩ or more :  1.00+0.02

Max.  

1μA

1000MΩ

OFF

 1200.0   MΩ 100 kΩ

100.0MΩ or less :  1.00+0.02

100.1MΩ or more : 10.00+0.02

*1

  

For negative values, to -10% f.s. The maximum display range is 9,999,999 dgt. or 9 GΩ. (An over-range error will be indicated when the maximum measurement range is exceeded, even 

if the maximum display range is not exceeded.)

*2 Measurement accuracy figures reflect accuracy after zero-adjustment. If not performing zero-adjustment, add the figures shown in the “Additional accuracy without 0ADJ” column. 

Figures shown in parentheses on the second line indicate the additional accuracy with OVC on.

*3 Measurement current accuracy is ±5%.

*4  When using an external trigger source or performing measurement with continuous measurement set to off (other than free-run), the open-circuit voltage from 1 ms after the completion 

of measurement (INDEX = ON) to the start of the next measurement (TRIG = ON) is limited to 20 mV or less.

  LP ON

    

• f.s. = calculated 100,000 dgt., where 0.001% f.s. = 1 dgt.

Range

100MΩ 

range high-

precision 

mode

Max.  

measurement  

display 

*1

Reso-

lution

Accuracy   %rdg. + %f.s. 

*2

Measurement 

current 

*3

Additional 

accuracy 

without 

0ADJ

%f.s. 

*2

Max open-

terminal 

voltage

FAST

MED

SLOW1

SLOW2

Switching

1000mΩ

  1200.00   mΩ

10 μΩ

0.200+0.010

0.200+0.010

0.200+0.005

0.200+0.003

1mA

20mV 

*5

10Ω

  12.000 0   Ω

100 μΩ

0.200+0.005

0.200+0.005

0.200+0.003

0.200+0.002

500μA

100Ω

  120.000   Ω

1 mΩ

0.200+0.005

0.200+0.005

0.200+0.003

0.200+0.002

50μA

1000Ω

  1200.00   Ω

10 mΩ

0.200+0.005

0.200+0.005

0.200+0.003

0.200+0.002

5μA

*1  For negative values, to -10% f.s. The maximum display range is 9,999,999 dgt. or 9 GΩ. (An over-range error will be indicated when the maximum measurement range is exceeded, even 

if the maximum display range is not exceeded.)

*2 Measurement accuracy figures reflect accuracy after zero-adjustment. LP values apply only when OVC is on.

*3 Measurement current accuracy is ±5%.

*5 When the contact check function is off (when the contact check function is on, 300 mV)

 

Resistance measurement accuracy

RM3545

RM3545

(

Example) 0.006 + 0.001 ....... 0.006

%

 rdg. + 0.001

%

 f.s.

Содержание RM3544

Страница 1: ...tion 0 01 LP 0 01m Basic accuracy 0 02 No of display digits Max 4 5 Max resolution 1 Featuring super high accuracy and multi channel capabilities 20 channels with 4 terminal measurement RESISTANCE MET...

Страница 2: ...or both manual operation and integration with automatic lines Max measurable current 1A Max resolution 0 01 Basic accuracy 0 006 Resistance measurement Max measurable current 1mA Max Open circuit volt...

Страница 3: ...on Pictured Back of the RM3545 02 with two Z3003 units optional feature installed Insert up to two Z3003 units into the slots on the back of the instrument Multi point measurement with the Multiplexer...

Страница 4: ...immediately after startup without any warm up or zero adjustment When performing measurement with the RM3545 in a temperature and humidity environ ment that satisfies the guaranteed accuracy condition...

Страница 5: ...act resistance indi cates an error Time R RM3545 Logger Auto scanning and step scanning When using the Multiplexer Unit Z3003 to perform scanning measure ment you can select either step scanning or au...

Страница 6: ...or saving and loading panels The RM3545 RM3544 can save and load up to 30 10 sets of range comparator and other settings Naming each set of panel data lets you make setup changes among production lots...

Страница 7: ...d to toggle the input signal polar ity between NPN sink output support and PNP source output support settings depending on the PLC common polarity EXT I O Input and Output Circuits EXT I O polarity Se...

Страница 8: ...ternal I O comparator BCD BIN function 3 3 3 Communication interfaces RS 232C Printer USB 3 3 3 GP IB N A 3 N A Multiplexer scanner function N A N A 3 Max 20 channels When using 4 terminal measurement...

Страница 9: ...rminal settings Front terminals MUX multiplexer When using the MUX setting the measurement leads cannot be connected to the front measurement terminals Support unit Z3003 Number of channels that can b...

Страница 10: ...0 m 0 008 0 005 0 007 0 002 0 007 0 001 1mA 20V 100k 120 000 0 k 100 m 0 008 0 005 0 007 0 002 0 007 0 001 100 A 1000k 1200 000 k 1 0 015 0 005 0 008 0 002 0 008 0 001 10 A 10M 12 000 00 M 10 0 030 0...

Страница 11: ...k 0 020 0 010 0 020 0 007 1mA 30k 35 000 k 0 020 0 010 0 020 0 007 100 A 300k 350 00 k 0 040 0 010 0 040 0 007 5 A 3M 3 500 0 M 0 200 0 010 0 200 0 007 500nA 6 For negative values to 10 f s 7 The maxi...

Страница 12: ...TTACHMENT L2105 2 m 6 56 ft RS 232C CABLE 9637 for PC connection 9pin 9pin cross 1 8 m 5 91 ft RS 232C CABLE 9638 for PC connection 9pin 25pin cross 1 8 m 5 91 ft GP IB CONNECTOR CABLE 9151 02 2 m 6 5...

Отзывы: