![Hioki PR8111 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/hioki/pr8111/pr8111_instruction-manual_2137342034.webp)
Chapter 4 Specifications
30
Recording paper
• Fanfold recording paper SE-10Z-2 (length: 15 m)
• Roll recording paper SE-10 (length: 20 m)
Maximum rated voltage
between terminals
250 VDC (V ranges) or 30 VDC (mV ranges)
Maximum rated voltage
between terminals and
ground
300 VAC, DC, measurement category II;
anticipated transient overvoltage: 2,500 V
(Between input channels and instrument or between input channels)
Dielectric strength and
insulation resistance
3.0 kVAC for 1 min, 50 M
Ω
or more at 500 VDC
(Between input channels and instrument or between input channels)
Battery check
3-stage LED display (green, orange, red) (when operating on battery power)
Accuracy is guaranteed even when red indicator is lit.
(green
→
orange: 6.9 V±0.15 V, orange
→
red: 6.4 V±0.15 V)
Operating temperature
and humidity
Temperature: 0°C to 40°C (32°F to 104°F)
Humidity: 40% to 80% RH (non-condensing)
Storage temperature and
humidity
Temperature: -20°C to 55°C (-4°F to 131°F)
Humidity: 10% to 80% RH (non-condensing)
Operating environment
Indoors, pollution degree 2, up to 2,000 m ASL
Dimensions
Approx. 292W × 177H × 182D mm (11.50”W × 6.97”H × 7.17”D)
(excluding protruding parts)
Mass
PR8111: Approx. 3.9 kg (137.6 oz) (instrument only)
PR8112: Approx. 4.4 kg (155.2 oz) (instrument only)
Power source
Model 9418-15 AC Adapter (12 VDC)
Rated supply voltage: 100 to 240 VAC (with allowance for 10% voltage fluctu-
ations)
Rated supply frequency: 50/60 Hz
Anticipated transient overvoltage: 2,500 V
DC power supply input: 10 to 27 VDC (Input can be provided to the AC adapter
jack via a special-order cable connected.)
Wiring between battery and the instrument must be 3 m or less in length.
Dry-cell batteries: Six LR20 alkaline batteries
(When used with the AC adapter, the adapter takes precedence.)
Max. rated power
4 VA (AC adapter, DC power supply) or 3 VA (dry-cell batteries)
Continuous operating time
When using LR20 alkaline batteries (reference values for 23°C)
PR8111: Approx. 50 hours
PR8112: Approx. 25 hours
Test conditions: 60 mm/h chart speed, ±250 mV measurement range, 400
mVp-p 0.01 Hz sine wave input waveform
Applicable standards
Safety: EN61010
EMC: EN61326 Class A
Effect of radiated radio-
frequency electromagnet-
ic field
±2% of effective recording with at 10 V/m
Effect of conducted radio-
frequency electromagnetic
field
±2% of effective recording width at 3 V
Содержание PR8111
Страница 1: ...PR8111 PR8112 Instruction Manual PEN RECORDER EN Feb 2020 Revised edition 3 PR8111A981 03 20 02H...
Страница 2: ......
Страница 4: ...Contents ii...
Страница 32: ...3 4 Example Printout 28...
Страница 36: ...Chapter 4 Specifications 32...
Страница 41: ...1...
Страница 42: ......
Страница 43: ......
Страница 44: ......