230
Waveform Calculation Settings
About Calculation Equations
Operators
Operators Name
Operators Name
ABS
Absolute Value
DIF2
2nd Derivative
EXP
Exponent
INT2
2nd Integral
LOG
Common Logarithm
SIN
Sine
SQR
Square Root
COS
Cosine
MOV
Moving Average
TAN
Tangent
SLI
Parallel displacement along the time axis
ASIN
Inverse Sine
DIF
1st Derivative
ACOS
Inverse Cosine
INT
1st Integral
ATAN
Inverse Tangent
PLCS
PLC delay shift
Refer to “11.3 Waveform Calculation Operators and Results” (p. 234).
To Enter Calculation Equations
•
Each calculation equation may contain up to 80 characters.
•
Each constant in a calculation equation may contain up to 30 digits.
• For multiplication, use [*]. For division, use [/].
• Entering a long, complex formula such as that shown below will cause
[?]
to be displayed. Separate the
formula into two shorter formulas.
ABS(CH1)+CH2*CH3-(CH4+CH5)*AB
1
2
3
•
If a formula tries to divide by 0, an overflow value will be output.
(For positive values, +9.9999E+29. For negative values, −9.9999E+29.)
• The result of calculation Z
i
can be used in other calculation equations. However, the Z
n
th equation can
only refer to the results of equations up to Z
n−1
.
(Example: Equation Z
4
can include the results of equations Z
1
through Z
3
.)
Using the MOV, SLI, DIF and DIF2 operators in an equation (The second parameter)
Set a number after the comma within the parenthesis (_,#) for each of these calculation operators.
Operators
Settings
Setting Examples
MOV
(Moving
Average)
SLI
(Horizontal
Movement)
Set the number of moving points to average over.
Setting Range
MOV (Moving Average): 1 to 5000
SLI: −5000 to 5000
Calculate the 10-point moving average of CH1:
MOV(CH1,10)
DIF
(Derivative)
DIF2
(2nd
Derivative)
Specify the sampling interval over which to
calculate the derivative.
“1” is normally acceptable, but this should be
set larger to capture fluctuation values of slowly
changing waveforms.
DIF and DIF2 Setting Range: 1 to 5000
Calculate the derivative of CH2 using a
20-point sampling interval:
DIF(CH2,20)
When calculation results overflow (OVER)
• The value at the AB cursor is not a correct value.
• When
[Scale]
is set to
[Auto]
, waveforms appear at the top or bottom edge of the screen. This makes
calculation result overflow obvious.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......