25
―――――――――――――――――――――――――――
Chapter 3 Specifications
――――――――――――――――――――――――
Open-circuit voltage
(when no load is applied)
1 to 1.2 times of rated
measurement voltage (open
circuit terminal voltage)
Lower limit measurement
resistance value to be
maintained rated
measurement voltage (M
Ω
)
0.25
0.5
2
Rated measurement current
(mA)
(current to be maintained
rated measurement voltage)
1 - 1.2
1 - 1.2
0.5 - 0.6
Short circuit current
(mA)
1.2 max
1.2 max 0.6 max
AC voltage measurement
AC voltage scale
(50/60 Hz) and allowance
0 to 500 V,
±
5%(=
±
25 V) of
maximum scale value
Input resistance
150 k
Ω
min
Maximum rated voltage
to earth
500 V AC
Input error protection for
10 seconds
1200 V AC
(overvoltage protection)