23
―――――――――――――――――――――――――――――――――――――――
Chapter 5 3184-02 GP-IB Specifications
―――――――――――――――――――――――――――――――――――
Bit 7
6
5
4
3
2
1
0
0
SRQ
0
0
0
SYNTAX
ERROR
OVER A OVER V
Status Byte
Service request sent
Undefined code, or code
containing over 21 characters
received
Overcurrent input
Overvoltage input
Binary
Decimal
Overvoltage input
01000001
65
Overcurrent input
01000010
66
Undefined code received
01000100
68
,
, and
occurring together
01000111
71
5.6 Service Requests
When the "S1" mode is set, a service request will be sent
in the event of an overvoltage or overcurrent input, or for
programming errors such as undefined codes or format
errors.
Following transmission of the service request, the
controller sends a serial poll, and the device responds by
transmitting its status byte and resetting the SRQ bit (bit
6).
When "S0" is set, a service request is not sent, but the
status byte is transmitted.
The OVER V or OVER A bit (bit 0 and 1) will be reset
with the serial poll, but a SYNTAX ERROR bit (bit 2)
remains until the next time the device is designated as
the listener, at which time it is reset.