62
Error
Error State
Solution
LEDs flash
quickly
three times.
LED is
lit up
continuously.
This issue occurs when the
DEMAG/
AUTO ZERO
key is pressed under the
circumstances described in the table
below.
Demagnetization and automatic zero-
adjustment cannot be performed.
Perform demagnetization and zero-
adjustment (p. 37) again after
implementing the remedy described
below.
Circumstances
Remedy
When the
JAW
UNLOCKED
LED is lit up
Lock the sensor
head.
(Press the opening
lever until the
JAW UNLOCKED
indicator turns off.)
When the
OVERLOAD
LED is flashing
Remove the
device from the
conductor being
measured.
When a current
of 0.5 A rms
or greater has
been detected
All LEDs
off even
though the
device is
receiving
power
The device has malfunctioned. Have it
repaired.