57
索
引
6
5
4
3
2
1
4
Repairs, Inspections, and Cleaning
4.1 Troubleshooting
Symptom
9HUL¿FDWLRQDQGRU6ROXWLRQ
The instrument is indicating an
abnormal measured value.
Is the measured current value too small for the instrument’s
measurement range?
Wrap the wire around the jaw one or more times. Each additional
wrap of the wire will increase the measured value, so that wrapping
it once yields a measured value that is twice the actual value and
wrapping it twice yields a measured value that is three times the
actual value.
Is the tip of the jaw open?
Is the jaw damaged?
If the jaw is damaged or cracked, it will not be able to measure current
accurately. Send the instrument for repair.
Displayed values can frequently
ÀXFWXDWHGXHWRLQGXFWLRQSRWHQWLDO
even when no voltage is applied. This, however, is not a malfunction.