2.5 Connecting the 9322 DIFFERENTIAL PROBE
______________________________________________________________
DANGER
For the 9322, the maximum rated voltage to earth is as
follows.
When using grabber clips: 1500 V AC or DC (CAT ll) /
600 V AC or DC (CAT lll)
When using alligator clips: 1000 V AC or DC (CAT lI) /
600 V AC or DC (CAT lll)
To avoid electrical shock and possible damage to the
product, never apply voltage exceeding these limits
between the input terminals and the chassis, or
between the inputs of two 9322s.
The maximum input voltage is 1000 V AC or 2000 V DC
(CAT ll) / 600 V AC or DC (CAT lll).
Do not attempt to measure voltage exceeding these
limits, as doing so may damage the equipment or
cause an accident that might result in injury or death.
WARNING
High voltage can be present at the inputs to the 9322.
To avoid electrical hazard, be sure to turn off the
equipment to be measured before connecting the test
leads. Turn the power on only after the leads have been
securely connected.
Before using the product, make sure that the sheathing
on the probes is not damaged and that no bare wire is
exposed. If there is damage, using the product could
cause electric shock. Contact your dealer or HIOKI
representative.