168
_____________________________________________________________________
8.11 Trigger Output Terminal
______________________________________________________________
Trigger output terminal
GND terminal
Output voltage
Open-collector signal (with voltage output),
active Low
Output voltage range
High level: 4.0 to 5.0 V
Low level: 0 to 0.5 V
Pulse width
Low level: Min. 30 ms
Recommended wire
Single strand: 1.0 mm dia. (AWG #18)
Multi-strand: 0.75 mm
2
Usable limits
Single strand: 0.3 to 1.0 mm dia.
(AWG #26 to #18)
Multi-strand: 0.3 to 0.75 mm
2
(AWG #22 to #20)
Strand diameter: Min. 0.18 mm dia
Standard insulation
stripping length
10 mm
Maximum input voltage
-20 V to +30 V
500 mA max, 200 mW max
10 k
Ω
5 V
TRIG.OUT
4.7 k
Ω
4.7 k
Ω
HIGH
4.0 to 5.0V
LOW
0 to 0.5V
30 ms or more
GND
8.11 Trigger Output Terminal
When triggering occurs, a signal is output from the TRIG
OUT terminal.
This can be used to synchronize several products.