Settings
______________________________________________________________
50 mV range
500 mV range
5 V range
50 V range
The preheat time cannot be defined to be longer than the
recording interval. Conversely, in the setup of the
recording interval, you cannot select a shorter recording
interval than the preheat time that is currently set.
If you choose a longer preheat time than the preset
recording interval and press the [Recording Interval] button
to complete the setup, the preset recording interval will
automatically be rewritten with a longer time than the
preheat time you have just chosen.
When you press the [Recording Interval] button to complete
the setup after setting a preheat time, the system will briefly
display the length of the recording interval that is currently
set. At this time you should verify the settings you have
made.
(3) Range setting
Pressing the Range button on the Monitor screen switches
the measurement range among 50 mV, 500 mV, 5 V, and
50 V.
As the measurement range is changed, the decimal point
will also change position.