4.2 Accuracy
46
Example accuracy calculation
Measurement range: 420.0 V AC
Accuracy specifications: ±1.2% (rdg.) ±3 (dgt.)
Measured value: 100.0 V
Since the value being measured is 100.0 V,
(A) Reading error ( % rdg.): 1.2% of 100.0 V = 1.2 V
(B) Digit error (dgt.): Maximum resolution of 0.1 V, so 3dgt. = 0.3 V
4.2 Accuracy
Accuracy guarantee for
temperature and hu-
midity
Guaranteed accuracy
period
23±5°C(73±9°F), 80%RH or less
1 year
Regulated power sup-
ply range
3.4 V or lower (until the
mark
lights up)
(rdg.: displayed value, dgt.: resolution)
Voltage Measurement
Range
[V]
Accuracy
±(rdg.)±(dgt.)
Input Impedance
(Frequency range)
V
420.0 m
4.200
42.00
420.0
1000
±0.5%±2
±0.5%±2
±0.5%±2
±0.5%±2
±0.5%±2
100 M
or more
Approx. 11M
Approx. 10M
Approx. 10M
Approx. 10M
V
*1
420.0 m
4.200
42.00
420.0
1000
±1.5%±3*2
±1.2%±3
±1.2%±3
±1.2%±3
±1.2%±6
100M
or more
(50 to 100 Hz)
Approx. 11M
(50 to 500 Hz)
Approx. 10M
(50 to 500 Hz)
Approx. 10M
(50 to 500 Hz)
Approx. 10M
(50 to 500 Hz)
Overload protection (for one minute):
1000 VDC, 1000 Vrms(sin), or 10
7
V
•
Hz
*1: The above measurement accuracies are applied to input of
at least 10% of the full scale.
Crest factor: 3 or lower (except 420.0 mV range)
*2: The above measurement accuracy for the 420.0 mV range
is applied to sine wave inputs.