27
2.4 Current Measurement
mA Measurement (40 mA/400 mA range)
Black
Red
REC•MEMO
1.
Move the function selector to the
mA
position.
2.
Select DC (
) or AC (
) using
the
REC•MEMO
button.
3.
Connect the red test lead to ter-
minal
A.mA
, and the black test
lead to terminal
COM
.
4.
Connect the test leads to the
measurement object, and read
the indicated value.
A Measurement (10 A range)
Note that the product may be damaged if current
exceeding the selected measurement range is
applied for a long time (for the 10 A range, continu-
ous current must be limited to 7A, or to less than
one minute if over 7A)
Black
Red
REC•MEMO
1.
Move the function selector to the
10A
position.
2.
Select DC (
) or AC (
) using
the
REC•MEMO
button.
3.
Connect the red test lead to ter-
minal
A
, and the black test lead
to terminal
COM
.
4.
Connect the test leads to the
measurement object, and read
the indicated value.
The Auto Range function is not
provided.