37
―――――――――――――――――――――――――――
Chapter 5Transfer Function
――――――――――――――――――――――――
NOTE
Item
Specifications
Remarks
Output data
Temperature measurement
data (fixed-length binary data)
Transmission
interval
0.2 s
Interface
RS-232C
No flow control
Signal
SD
、
GND
Transmission only;
no reception
Transfer format
Asynchronous : 38400 bps
Data bit : 8
Start bit : 1
Stop bit : 2
Parity bit : None
Chapter 5
Transfer Function
5.1 Interface Specifications
Temperature measurement data is sent from the
interface terminal on the main unit. Temperatures
can be recorded, monitored, and alarms sent by
connecting to a PC and creating a program.
The data transfer function on the main unit is only
able to transmit temperature measurement data
repeatedly. The unit cannot be controlled via a PC
(e.g. to change settings) and it is not possible to
acquire camera image data.