194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
Page 47
Mob
ili
zi
n
g
th
e pat
ien
t
Display the weight to within 100 g
1.
Press
on the GCI.
2.
Press the command
3.
The weight is displayed to within 100 g
for 5 seconds.
Add/remove parts on the bed
Caregivers can use this function to change the parts on the bed without changing the
weight reading on display.
If a patient is
not
on the bed, use the Zero/Tare function after changing the parts.
The function stores the patient’s weight in memory
while changing items on the bed.
1.
Press
on the home screen of the GCI.
2.
Press
. Follow the
instructions.
The list of items added or removed, displayed
close to the bed, may be useful later on.
Weighed/unweighed parts
The weigh system is very sensitive. The weight reading is more precise if the part of the
bed that is weighed (sleep surface, siderails, footboard – see weighed parts below) does
not come into contact with the fixed part of the bed (headboard, chassis, power supply
cables and tubes of the various possible accessories – see unweighed parts below). Even if
an object is in slight contact with the bed, the value on the screen will be inaccurate. If
necessary, remove the headboard during scale.
The screen shows the zone and the parts of the bed that are included when weighing in
green.
Содержание Accella 900
Страница 1: ... 194412 EN Hill Rom 900 Accella Bed Instructions for Use LI900B4 194412 Rev 11 ...
Страница 6: ...Page iv 194412 11 Hill Rom 900 Accella Bed Instructions For Use ...
Страница 26: ...Page 20 194412 11 Hill Rom 900 Accella Bed Instructions For Use ...
Страница 80: ...Page 74 194412 11 Hill Rom 900 Accella Bed Instructions For Use ...
Страница 102: ...Page 96 194412 11 Hill Rom 900 Accella Bed Instructions For Use ...