![Hillrom 900 Series Скачать руководство пользователя страница 81](http://html.mh-extra.com/html/hillrom/900-series/900-series_instructions-for-use-manual_2134308081.webp)
171051(12) - Hillrom™ 900 Electric Bed Instructions for Use
Page 75
De
contamina
tio
n
M
aintenanc
e
Decontamination, Maintenance
Decontamination of the bed
Safety recommendations
• Ensure that the bed cannot move.
• Lock out all electrical functions.
• Disconnect the bed and stow the power cable (see “Securing the power
• Check that all plugs are well connected (control and lockout units, electric
motors on the power supply unit).
• Never clean the bed by pouring water on it, nor with high-pressure hoses nor
in tunnel washes.
• Never use water at a temperature of more than 60°C.
• Avoid excess water on the connectors.
• Refer to the recommendations of the cleaning product manufacturer.
• Thoroughly dry before reusing.
Failure to implement one or more of these recommendations may lead to damage
or deterioration, preventing use of the bed and rendering the warranty void.
Recommendations for cleaning and disinfection
The following recommendations are not designed to replace existing cleaning
protocols drawn up by the hygiene officer or by other bodies for your hospital.
The disinfecting method described below applies specifically to the bed and its
accessories and is designed to save time and to help combat nosocomial
infection more effectively.
Clean the bed with a lightly dampened cloth and ordinary disinfectant. Do not use
excessive liquid.
This bed is designed for easy cleaning and optimal hygiene.