4.10 Electric (P8020) Stretcher—Patient Control Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.10 Electric (P8020) Stretcher—Patient Control Switch
Tools required:
head bit
Ratchet
Removal
SHOCK HAZARD:
Failure to unplug the stretcher from its power source could cause
personal injury or equipment damage.
1.
Unplug the stretcher from its power source.
2.
Remove the two screws (D) that attach the cover housing (C) under the
patient control unit (A) (see figure 4-13 on page 4-27).
3.
Remove the cover housing (C), and disconnect the patient control switch
connector.
Figure 4-13. Patient Control Switch
4.
Remove the patient control switch (B) from the patient control unit (A).
1. Torx® is a registered trademark of Acument Intellectual Properties, LLC.
4
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 4 - 27
Service Manual (144386 REV 2)
Содержание P8000
Страница 2: ......
Страница 14: ...NOTES Page xii Hill Rom Transport Procedural and Specialty Stretchers Service Manual 144386 REV 2 ...
Страница 441: ...Schematic Wiring Diagram Control Board P N 49603 ...
Страница 442: ...Schematic Wiring Diagram Electric Stretcher P N 60389 ...
Страница 443: ......