Instructions for Use
DRAFT 29-NOV-2016
component, do
not
use it, and replace that component before
the next therapy session. Failure to do so could cause injury.
•
If the garment is wet or damp, do not plug it into a power source
or use it.
NOTE:
If your physician has prescribed aerosol therapy to be used during the
treatment, set up the appropriate equipment.
Getting Started
Install Battery
1.
Lift the battery access panel.
2.
Align the battery so the battery latch
is toward the top of the garment.
3.
Insert the bottom of the battery into
the battery panel first.
4.
Slide the latch then proceed to insert
the top of the battery until the
battery locks into place.
NOTE:
The images show how to confirm the battery is locked into place. If you
see an orange color in the latch area, then the battery is not fully locked
into place. Press down on the battery until the orange color is no longer
visible.
CAUTION:
Fully charge the battery to 100% before first time use.