2.2 Troublesshooting Viking™ M mobile lift, 2040015
1. Make sure that the emergency stop is not engaged.
2. If the lift has not been used for 24 hours, enable the system.
3. Check that the HandControl IR is enabled.
4. Check that the charger cable is not connected to an electric outlet.
5. HandControl IR indicates battery change.
6. If the lift works via the operation panel,
change the HandControl IR.
7. If the problem persists, please contact Hill-Rom.
1. Make sure that the emergency stop is not engaged.
2. Make sure that the cables to the control box are connected
correctly.
3. Check that the charger cable is not connected to an electric outlet.
4. Check the battery capacity.
5. Check that the battery’s contact plates are not defective or broken.
6.
If the problem persists, please contact Hill-Rom.
1. Make sure that the emergency stop is not engaged.
2. Check that the charging cable is correctly connected
3. Make sure that the battery is properly attached.
4. Check that the battery’s contact plates are not defective or broken.
5. If the problem persists, please contact Hill-Rom.
1. Make sure that the emergency stop is not engaged.
2. Check the battery capacity.
3. HandControl IR indicates battery change.
4. Use the electrical emergency lowering device to lower
the patient onto a firm surface.
5. Use the manual emergency lowering device to lower
the patient onto a firm surface.
6.
If the problem persists, please contact Hill-Rom.
Contact Hill-Rom.
The lift do not work up/down with the
Hand Control IR
The base-width adjustment doesn’t
work (in/out) with the Hand Control IR.
The lift does not work up/down with
the operation panel.
The base-width adjustment doesn’t
work (in/out) with the operation panel.
The charger doesn’t work.
The lift is stuck in the high position.
If you hear unusual sound from the lift.