High Voltage PFT Series Скачать руководство пользователя страница 21

Saved 10/7/2008 9:08:00 AM, T:\PRODUCT\DOCUMENTS\Manuals\PFT\pl_pft1003CM_2k8Ri.doc 

Page 1 of 3 

Parts List PFT-1003CM (120V) 

Schematic # PFT-1027S 

REF. QUAN HVI# DESCRIPTION 

 

 

 

 

 

CAB 

32-033 

CABINET, PORTABLE TRANSIT ENCLOSURE. CHARCOAL GRAY.18W x 10D x 14 H ID 

 

 

 

 

D1 1 

04-030 

DIODE, 

1N5408A 

D2,3 2 

04-025 

DIODE, 

1N4007A 

 

 

 

 

F1 

06-043 

FUSE, 15A, 250V, ABC-15 

 

06-015 

FUSE HOLDER, LITTELFUSE # 342004A 

 

 

 

 

J2 

07-440 

CONNECTOR, BULKHEAD, 2 SOCKET, CINCH #S302AB 

J3 

 

CONNECTOR, BULKHEAD, CIRCULAR, PART OF HIGH VOLTAGE TANK 

J4 1 

07-206 

CONNECTOR, 

BULKHEAD, BNC, AMPHENOL # UG1094A/U 

 

 

 

 

K1 

11-150 

CONTACTOR, 3PDT , 15A, 120 Vac COIL, MAGNECRAFT # W389ACX-14 

 

 

 

 

M1 1 

13-141 

METER, ANALOG, 100 

µ

A MOVEMENT, SCALED 0-50/100 AC KILOVOLTS 

M2 1 

13-151 

METER, ANALOG, 100 

µ

A MOVEMENT, SCALED 0-1.0 AC MILLIAMPERES 

 

 

 

 

MOV1 

06-207 

METAL OXIDE VARISTOR, # V130LA10A 

MOV2 

06-210 

METAL OXIDE VARISTOR, # V250LA20A 

 

 

 

 

P1 

22-205 

INPUT POWER CABLE, 16/3, 12 ft.  

P2 

07-442 

CONNECTOR, CABLE, 2 CIRCUIT, CINCH# P302CCT 

P3 

07-014 

CONNECTOR, CABLE, 7 PIN CIRCULAR, AMPHENOL # 97-3106B-20-15S 

 

 

 

 

PCB-001-PFT10 

82-275 

PRINTED CIRCUIT BOARD, VOLT METER  

PCB-034 

82-274 

PRINTED CIRCUIT BOARD, AC OVERLOAD/CURRENT METER  

 

 

 

 

S1 

10-214 

SWITCH, PB, MAINT’D, 250 Vac, 15A, DPST, NO , NEON LAMP, MICROSWITCH 
#AML32FBB4AD 

 

10-250 

SWITCH, PB COVER, BLACK, MICROSWITCH # AML52-N10K 

S2 

10-222 

SWITCH, PB, MOM., 125 Vac, 15A, SPDT, GRN LED, MICROSWITCH # AML22CBS2AA 

S3 

10-218 

SWITCH, PB, MOM., 125 Vac, 15A, SPDT, RED LED, MICROSWITCH # AML22CBC2AA 

 

10-252 

SWITCH, PB COVER, BLACK, MICROSWITCH # AML52-C10K 

S4 

10-106 

SWITCH, SNAP ACTION, SHORT ARM, OMRON #A-20GV22-B7-K 

S5 

10-514 

SWITCH, ROTARY, 6 POL, 2 POS, MOUSER # 10YX062 

S6 

10-510 

SWITCH, ROTARY, 4 POL, 3 POS, MOUSER # 10WW043 

S7 

10-402 

SWITCH, ROCKER, DPDT, CARLINGSWITCH # 62115929-0-0-V 

 

 

 

 

SG1-3 

06-205 

SPARK GAP, 90V, CP CLARE, CG90L 

 

 

 

 

T1 

25-108 

TRANSFORMER, VARIABLE, SUPERIOR TYPE 116 

T2 

T080 

TRANSFORMER, CURRENT, HVI #080 

 

 

 

 

HV TNK 

84-240 

TANK, HIGH VOLTAGE 100 kVac ,  HV # PFT-1008S REV C 

 

 

 

 

PCB-001-

PFT10 

 

82-275 

VOLT METER PCB 

C1 1  03-060 

CAPACITOR, ELECTROLYTIC, RADIAL LEADS, 100

µ

F, 50 Vdc, MOUSER # 140-XRL50V100 

D1 1  04-025 

DIODE, 

1N4007A 

D2 

04-415 

DIODE, FULL WAVE BRIDGE, 1.5A, 1kVdc, MOUSER # 583-RB157 

J1 

07-136 

CONNECTOR, HEADER, 8 PIN, .1” SPACING, MOLEX # 22-23-2081 

NE1,2 

15-114 

NEON LAMP, #NE2 

P1 

07-120 

CONNECTOR, CABLE, 8 PIN, .1” SPACING, MOLEX # 22-01-2087 

 

07-104 

CONNECTOR PINS, .1” SPACING, MOLEX # 08-50-0114 

PCB 

14-001 

PRINTED CIRCUIT BOARD, HVI # PCB- 001 

R1,2 

01-154 

RESISTOR, METAL FILM, 0.25W, 200K, 1% 

Содержание PFT Series

Страница 1: ... power source The operator of this equipment must use good judgement and follow all safety precautions noted in this guide to ensure the protection of himself and others in close proximity to the test area Failure to follow the instructions could result in injury or death Proper grounding of the test set must be done prior to connecting this unit to a power source PFT SERIES Please Refer to Docume...

Страница 2: ...OperatorManual ...

Страница 3: ... 1 General Information 2 Features and Specifications 2 3 Controls and Indicators 4 5 S E C T I O N 2 Setting up the Equipment 6 Operating the Equipment 7 10 Blank Page for Notes 11 S E C T I O N 3 Performing Special Operations 12 Meter Re calibration 12 13 Miscellaneous 14 Packing the Cables 14 Return Material 15 Warranty 16 17 ...

Страница 4: ... to established procedures for the use of this type of equipment This Operator Manual is organized to provide information on the PFT Series of Hipots in steps that familiarize the new operator with the operation of this test set Section 1 Specifications and Controls Section 2 Setup and Operation Section 3 Performing Special Operations TheFunctions Features andSpecifications ofthePFTSeriesof ACHipo...

Страница 5: ...urrents The new simplified controls incorporate this feature transparently Standard features of the PFT Series of include Continuously adjustable output voltage Fixed overload factory set to 120 of variable transformer rated output current Zero Start and External Interlock provision Secondary connected dual range voltmeter Shielded output cable 50 kV and below precludes unnecessary insulated suppo...

Страница 6: ...3 PFT MODEL SPECIFICATIONS PFT 103 PFT 303 PFT 503 PFT 1003 SEE SPECIFICATIONS ON LAST SHEET OF PARTS LIST Table 1 PFT Series Specifications ...

Страница 7: ...R N A L I N T E R L O C K The Ext Intlk connector is provided to allow for a normally open safety interlock switch to control the energizing of the high voltage output H I G H V O L T A G E O N O F F The HIGH VOLTAGE ON and OFF pushbuttons control the high voltage power circuits The red and green LEDs provide life long positive indication of the circuit status The red LED lights when high voltage ...

Страница 8: ... O D E The RETURN MODE switch allows for accurate secondary load current readings In the GUARD position all ground currents are shunted around the meter The load must be isolated from ground to use in the GUARD position In the GROUND position the return point is grounded and all currents load and ground losses are metered B U R N O P T I O N A L The BURN toggle is provided for tests when the load ...

Страница 9: ...en provided for the ground connection A second ground lead is provided for two piece units for the grounding of the high voltage section 4 Insert the EXT INTLK plug into the socket on the panel The plug may also be wired to a normally open contact of a safety switch for added protection Hand held safety switches are available from High Voltage to plug directly into the EXT INTLK socket 5 If applic...

Страница 10: ...METER RANGE for the test voltage level expected 3 Prior to connecting the output cable toroid to the test sample be sure that the test sample is de energized 4 Connect the output lead to the test sample Be sure that there is enough clearance to grounded objects for the expected test voltage The minimum clearance in air is 10 kV ac inch 5 Return the low side of the test sample to the LOAD RETURN co...

Страница 11: ...er the test is complete rotate the OUTPUT control to zero prior to depressing the HV OFF pushbutton 12 If the test sample fails during the test the internal overload relay will de energize the high voltage This relay is in the primary circuit and is sensitive to primary current overloads The overload is set to 120 of the rated current of the variable transformer The use of a gapped transformer in ...

Страница 12: ... 1 Grounded Return With the output return in the grounded mode the current meter reads all current to ground internal and external to the test set This current might include corona surface tracking and any shunt resistance The typical diagram for grounded return operation is shown in Figure 2 below 2 Guarded Return With the output return in the guarded mode the current meter will only read current...

Страница 13: ...rn Diagram Test Sample High Voltage From Test Set Current flow through return includes only current through the test sample Note that the current to ground is returned after the meter Meter GROUND GUARD RETURN POST GROUND POST ...

Страница 14: ...11 THIS PAGE LEFT BLANK FOR APPLICATION NOTES ...

Страница 15: ... the aging of the meter The certification of meters on a yearly basis is recommended to ensure accurate test results Voltmeter Re calibration 1 Locate the unit in a position that will allow easy reading of the meters 2 Remove the panel screws and support the panel vertically to gain access to the calibration pot on the back of the voltmeter 3 Zero the meter movement using the zero adjustment below...

Страница 16: ...lid earth ground using the supplied black ground test lead prior to connecting the unit to input power 14 Set the CURRENT RANGE to X1 position 15 Connect the output cable to a calibrated reference meter with ability to read 1mA ac Use a series resistance or capacitance to allow for precise resolution when calibrating Be sure to connect the low side of the reference current meter to LOAD RETURN and...

Страница 17: ...nk are available from HIGH VOLTAGE INC at the address noted on the inside front cover of this manual The oil level in the tank should be 5 inches from the lid when the oil temperature is 20 C Packing the Cables The output cable will easily fit into the cable storage area next to the control panel Care should be used to avoid damaging the output cable jacket when coiling into the compartment As the...

Страница 18: ...rn of the equipment is appropriate If deemed appropriate a Return Authorization Number will be issued At that time the Purchaser will be instructed how to mark and return the equipment The above procedure must be adhered to in order to ensure prompt service No equipment should be returned without the prior knowledge and authorization of High Voltage Inc REPLACEMENT PARTS ORDERING To order replacem...

Страница 19: ...red by the purchaser for labor and or materials incidental to the inspection repair replacement or issuance of credit for product or materials under warranty is the sole responsibility of the purchaser This warranty applies solely to products manufactured by High Voltage It does not apply to parts accessories or materials not manufactured by High Voltage or to consumables If purchaser s claim rela...

Страница 20: ...lusively determined by High Voltage and any determination shall be final and binding THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED ON THE PART OF HIGH VOLTAGE INCLUDING THE WARRANTIES OF MERCHANTABILITY FITNESS FOR A PARTICULAR USE AND NON INFRINGEMENT ARISING FROM ANY BREACH THEREOF AND HIGH VOLTAGE NEITHER ASSUMES NOR AUTHORIZES ANY OTHER PERSON FIRM OR CORPORAT...

Страница 21: ... AC OVERLOAD CURRENT METER S1 1 10 214 SWITCH PB MAINT D 250 Vac 15A DPST NO NEON LAMP MICROSWITCH AML32FBB4AD 1 10 250 SWITCH PB COVER BLACK MICROSWITCH AML52 N10K S2 1 10 222 SWITCH PB MOM 125 Vac 15A SPDT GRN LED MICROSWITCH AML22CBS2AA S3 1 10 218 SWITCH PB MOM 125 Vac 15A SPDT RED LED MICROSWITCH AML22CBC2AA 2 10 252 SWITCH PB COVER BLACK MICROSWITCH AML52 C10K S4 1 10 106 SWITCH SNAP ACTION ...

Страница 22: ...Ω 1 R4 7 2 02 018 RESISTOR POTENTIOMETER 0 25W 10K MOUSER 569 25PR 10K R5 1 01 098 RESISTOR METAL FILM 0 25W 7 68K 1 R6 1 01 138 RESISTOR METAL FILM 0 25W 49 9K 1 R8 9 2 01 312 RESISTOR CARBON FILM 1W 1 6K OHM 5 RY1 2 2 11 140 RELAY PCB MOUNT SPDT 24 Vdc COIL P B T70L5D131 24 MISC 1 23 101 KNOB SKIRTED DIAL MOUSER 45KN021 2 23 101 KNOB SKIRTED POINTER ALCO PKD 70B 1 4 1 BRACKET ZERO START HVI BKT ...

Страница 23: ...tive load 3KVA capacitive load Output Termination Top Toroid Duty 3KVA 1 hour ON 1 hour OFF 1 5KVA continuous Distortion 5 Meter Accuracy 2 FS Kilovoltmeter 3 5 inch Scaled 0 50 100kVac RMS Current Meter 3 5 inch Scaled 0 1 0 milliamperes with x1 x10 x100 range multiplier Case Size H V TANK 19w x 11d x 15 5 high 12w x 13d x 24 high Weight CONTROL 50 lbs 23kg HV TANK 82 lbs 38kg Table 1 PFT 1003CM ...

Страница 24: ...D AC OVERLOAD CURRENT METER S1 1 10 214 SWITCH PB MAINT D 250 Vac 10A DPST NO ISOLATED NEON LAMP MICROSWITCH AML32FBB4AD 1 10 250 SWITCH PB COVER BLACK MICROSWITCH AML52 N10K S2 1 10 222 SWITCH PB MOM 125 Vac 15A SPDT GRN LED MICROSWITCH AML22CBS2AA S3 1 10 218 SWITCH PB MOM 125 Vac 15A SPDT RED LED MICROSWITCH AML22CBC2AA 2 10 252 SWITCH PB COVER BLACK MICROSWITCH AML52 C10K S4 1 10 106 SWITCH SN...

Страница 25: ...SISTOR METAL FILM 0 5W 133Ω 1 R4 7 2 02 018 RESISTOR POTENTIOMETER 0 25W 10K MOUSER 569 25PR 10K R5 1 01 098 RESISTOR METAL FILM 0 25W 7 68K 1 R6 1 01 138 RESISTOR METAL FILM 0 25W 49 9K 1 R8 9 2 01 312 RESISTOR CARBON FILM 1W 1 6K OHM 5 RY1 2 2 11 140 RELAY PCB MOUNT SPDT 24 Vdc COIL P B T70L5D131 24 MISC 1 23 101 KNOB SKIRTED DIAL MOUSER 45KN021 2 23 101 KNOB SKIRTED POINTER ALCO PKD 70B 1 4 1 B...

Страница 26: ...istive load 3KVA capacitive load Output Termination Top Toroid Duty 3KVA 1 hour ON 1 hour OFF 1 5KVA continuous Distortion 5 Meter Accuracy 2 Kilovoltmeter 3 5 inch Scaled 0 50 100kVac RMS Current Meter 3 5 inch Scaled 0 1 0 milliamperes with x1 x10 x100 range multiplier Case Size H V TANK 19w x 11d x 15 5 high 12w x 13d x 24 high Weight CONTROL 55 lbs 25kg HV TANK 82 lbs 38kg Table 1 PFT 1003CMF ...