Global sphere measurement
Digitize the whole sphere with the scanner: wait for the 100 % coverage at least to stop digitizing.
Validate by pressing on BP2 (right button).
It is still possible to continue scanning even once the covered is completed, to
enhance the quality of calibration, however, the calculation may be longer.
Calculation
After all the measurements, the calculation automatically starts.
At the end of calculation, click on yes.
Verification
10. To validate the calibration, please perform the laser scanner accuracy test : digitize the sphere
,then stop and check the results.
For more details, please refer to RDS Data Collector manual.
Probe calibration is finished.
Residual deviation=0.019797
Do you wish to use calculated calibration ?
Please wait.
Step=9, Residual error=0.019797, Improvement=0.005339%
Cancel
Yes
No