18/34
5 Settings and enquiries
Rotation area
RPM x 100
START
STOP
t
Speed area
Time area
Keypad
5.1
Setting the machine version
The machine version stored in the control panel must correspond to the centrifuge
model. Control panels supplied as spare part are already set to machine version 4
(ROTOFIX 32).
If another machine version than 4 is set in the control panel, the set machine version will
appear in the display after switching on the centrifuge ROTOFIX 32.
The machine version 4 must be set as follows:
1. Adjust the machine version 4 by
using the and arrow keys in
the time area.
RPM x 100
t
Maschinenversion / Machine version
START
STOP
2. Press the key
STOP
in order to store the set machine version.
5.2
Enquiry the machine version
1. Switch off the mains switch.
2. Simultaneously press the pulse key and the arrow key in the speed area.
Switch on the mains switch and release the keys again.
3. If necessary press the key so often until the machine version appears in the
speed area and the brake setting appears in the time area.
RPM x 100
START
STO P
t
Speed area Time area
RPM x 100
START
STO P
t
Speed area Time area
4. To exit the machine version display press the key
STOP
or perform a MAINS-
RESET.