EN
AB2300DEENFRIT
Rev. 07
31/66
13.1 Centrifugation
with preselected time
Set the desired speed with the keys beneath the speed indicator.
Set the desired time with the keys beneath the time indicator.
Press the key
START
. The rotation indicator appears while the rotor is turning.
The time is displayed in minutes. The last minute is counted down in seconds.
When the time is displayed in minutes, a point flashes next to the number.
After expiry of the time or if the centrifugation run is interrupted by pressing the key
STOP
, the rotor runs down
with the set brake step.
During the centrifugation run the rotor speed and the remaining time are displayed.
13.2 Continuous
operation
Set the desired speed with the keys beneath the speed indicator.
Set the time to zero with the key beneath the time indicator. "
--
" is displayed.
Press the key
START
. The rotation indicator appears while the rotor is turning. The time count starts from 0.
The first minute is counted up in seconds, and then the time is displayed in minutes.
When the time is displayed in minutes, a point flashes next to the number.
Press the key
STOP
to end the centrifugation run. The rotor runs down with the set brake step.
During the centrifugation run the rotor speed and the expired time are displayed.
13.3 Short-time
centrifugation
Set the desired speed with the keys beneath the speed indicator.
Keep the key
PULSE
pressed. The rotation indicator appears while the rotor is turning. The time count starts
from 0.
The first minute is counted up in seconds, and then the time is displayed in minutes.
When the time is displayed in minutes, a point flashes next to the number.
Release the key
PULSE
again to end the centrifugation run. The rotor runs down with the set brake step.
During the centrifugation run the rotor speed and the expired time are displayed.
14
Relative centrifugal force (RCF)
The relative centrifugal force (RCF) is given as a multiple of the acceleration of gravity (g). It is a unit-free value and
serves to compare the separation and sedimentation performance.
These values are calculated using the formula below:
1,118
r
2
1000
RPM
RCF
1000
1,118
r
RCF
RPM
RCF = relative centrifugal force
RPM = rotational speed (revolutions per minute)
r = centrifugal radius in mm = distance from the centre of the turning axis to the bottom of the centrifuge
container. For more on the centrifugal radius see the chapter
“Anhang/Appendix,Rotoren und Zubehör/Rotors and accessories”.
The relative centrifugal force (RCF) stands in relation to the revolutions per minute and the centrifugal
radius.