Vet/IV 2.2
Infusion Pump
14
Section 10: Special Programs
10.1 Select Infusion Modes and Pressure Settings
1. With the pump off, press and hold
SELECT
.
2. Press
ON/OFF
quickly
then release.
3. [SLCt] shows on the display. Release
SELECT
.
4. Press
SELECT
to scroll through the available infusion modes. The following modes are available:
MACRO
MIC (Micro)
10-P
TPN
TPN Easy
5. When the desired mode is illuminated on the right side of the display, press
RUN/HOLD
to store the
selection in memory.
6. The display changes to [PrES] followed by the last selected pressure setting [HI, Lo, nor]. Press
SELECT
to scroll between High, Low, or normal infusion pressures.
NOTE:
This setting determines the pressure point at which the occlusion alarm is sounded. See
Pressure
Settings Definitions
in this section for details. Rate, volume,
etc
. are not affected by this setting.
NOTE:
The HI pressure setting is recommended for almost all conditions, to minimize unnecessary
alarming.
7. When the desired pressure setting is displayed, press
RUN/HOLD
to store the setting in memory.
8. Continue programming Rate/Time or Rate/VTBI as usual.
10.2 Pressure Settings Definitions
Low = 0.4 bar (6 psi) – for geriatrics, neonates, patients with fragile veins
Normal = 0.6 bar (8 psi)
High = 0.8 bar (12 psi)
Recommended setting for most patients.
NOTE:
The pump recalls programmed pressure settings indefinitely until they are changed by the operator. Once new
pressure settings are stored in any program, previous settings in all programs are erased.
10.3 10-P Program
1. Set up the pump as usual, choosing 10-P as the infusion mode.
2. The display shows [P-1], followed quickly by [0] (zero). Rate will be illuminated on the left side of the
display. This indicates that you will be setting the first rate/volume combination in the 10-program
series, starting with the first infusion rate.
3. Set Rate and VTBI as previously discussed, pressing
SELECT
after each entry.
NOTE:
Use
SELECT
to move between parameters and enter settings.
Do not press RUN/HOLD
until you
have entered all desired programs and are ready to begin infusion.