Vet/IV 2.2
Infusion Pump
10
Section 7: Program Descriptions
7.1 Programs
The Vet/IV Pump features 6 different programs which can be used for a variety of applications. Following are the
specifications of each program.
NOTE:
The pump will continuously display the program setting by highlighting the appropriate indicator on the
right side of the display.
7.1.1 MACRO
MACRO
Macro program.
Rate: 1 to 999 ml/hr in 1 ml increments up to 100 ml/hr, then in 10 ml increments up to 999 ml/hr.
Total volume to be infused: 1 to 9,999 ml.
NOTE:
Standard IV sets are recommended for use with this mode.
7.1.2 MICRO
MIC
Micro program for smaller animals or micro infusions.
Rate: 0.1 to 99.9 ml/hr in 0.1 ml increments up to 10 ml/hr, then in 1 ml increments up to 99.9 ml/hr.
Total volume to be infused: 0.1 to 2,500 ml.
NOTE:
Micro IV sets are recommended for use with this mode.
7.1.3
10-P Multi-Program
10-P
Special program which enables the user to program up to 10 different rate/volume combinations
for a single patient. Each program starts immediately after completing the previous program. Note
that you do not have to enter 10 different combinations to use this program. You may enter as few as
1 or as many as 10 programs, as needed for your patient.
Each program can be set with the following parameters:
Rate: 1 to 999 ml/hr.
Total volume to be infused: 1 to 9,999 ml.
NOTE:
Total volume of all programs cannot exceed 9,999 ml/hr.
7.1.4 TPN
TPN
For administration of Total Parenteral Nutrition. Allows user to set the up and down ramping time
and total time of infusion.
7.1.5
TPN EASY
Delivers TPN with automatic 1-hour ramp-up and 1-hour ramp-down times.
Volume To Be Infused and total Time of Infusion are set by user.
Pump calculates Maximal Rate.
7.1.6
PIGY Program
Delivers fluids using two different sequential rates and volumes to accommodate piggy-back
administration systems.
NOTE:
Some mixing of fluids and/or back-flow of fluid from the main bag into the supplemental bag
may occur with the use of this program. This is a recognized phenomenon in fluid therapy and has
been noted with both pump administration and gravity feed systems. For this reason, do not use the
PIGY function to administer chemotherapeutic agents or other solutions that should not be mixed.